An aberration correction approach for single and dual aperture ultrasound imaging of the abdomen.

Ultrasonics

Photoacoustics & Ultrasound Laboratory Eindhoven (PULS/e), Department of Biomedical Engineering, Eindhoven University of Technology, P.O. Box 513, Eindhoven, 5600 MB Eindhoven, The Netherlands. Electronic address:

Published: May 2023

Abdominal ultrasound image quality is hampered by phase aberration, that is mainly caused by the large speed-of-sound (SoS) differences between fat and muscle tissue in the abdominal wall. The mismatch between the assumed and actual SoS distribution introduces general blurring of the ultrasound images, and acoustic refraction can lead to geometric distortion of the imaged features. Large aperture imaging or dual-transducer imaging can improve abdominal imaging at deep locations by providing increased contrast and resolution. However, aberration effects for large aperture imaging can be even more severe, which limits its full potential. In this study, a model-based aberration correction method for arbitrary acquisition schemes is introduced for delay-and-sum (DAS) beamforming and its performance was analyzed for both single- and dual-transducer ultrasound imaging. The method employs aberration corrected wavefront arrival times, using manually assigned local SoS values. Two wavefront models were compared. The first model is based on a straight ray approximation, and the second model on the Eikonal equation, which is solved by a multi-stencils fast marching method. Their accuracy for abdominal imaging was evaluated in acoustic simulations and phantom experiments involving tissue-mimicking and porcine material with large SoS contrast (∼100 m/s). The lateral resolution was improved by up to 90% in simulations and up to 65% in experiments compared to standard DAS, in which the use of Eikonal beamforming generally outperformed straight ray beamforming. Moreover, geometric distortions were mitigated in multi-aperture imaging, leading to a reduction in position error of around 80%. A study on the sensitivity of the aberration correction to shape and SoS of aberrating layers was performed, showing that even with imperfect segmentations or SoS values, aberration correction still outperforms standard DAS.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultras.2023.106936DOI Listing

Publication Analysis

Top Keywords

aberration correction
16
imaging
8
ultrasound imaging
8
large aperture
8
aperture imaging
8
abdominal imaging
8
sos values
8
straight ray
8
standard das
8
aberration
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!