Rationale And Objectives: In breast MRI with diffusion-weighted imaging (DWI), fat suppression is essential for eliminating the dominant lipid signal. This investigation evaluates a combined water-excitation-spectral-fatsat method (WEXfs) versus standard spectral attenuated inversion recovery (SPAIR) in high-resolution 3-Tesla breast MRI.
Materials And Methods: Multiparametric breast MRI with 2 echo-planar DWI sequences was performed in 83 patients (50.1 ± 12.6 years) employing either WEXfs or SPAIR for fat signal suppression. Three radiologists assessed overall DWI quality and delineability of 88 focal lesions (28 malignant, 60 benign) on images with b values of 800 and 1600 s/mm, as well as apparent diffusion coefficient (ADC) maps. For each fat suppression method and b value, the longest lesion diameter was determined in addition to measuring the signal intensity in DWI and ADC value in standardized regions of interest.
Results: Regardless of b values, image quality (all p < 0.001) and lesion delineability (all p ≤ 0.003) with WEXfs-DWI were deemed superior compared to SPAIR-DWI in benign and malignant lesions. Irrespective of lesion characterization, WEXfs-DWI provided superior signal-to-noise, contrast-to-noise and signal-intensity ratios with 1600 s/mm (all p ≤ 0.05). The lesion size difference between contrast-enhanced T1 subtraction images and DWI was smaller for WEXfs compared to SPAIR fat suppression (all p ≤ 0.007). The mean ADC value in malignant lesions was lower for WEXfs-DWI (p < 0.001), while no significant ADC difference was ascertained between both techniques in benign lesions (p = 0.947).
Conclusion: WEXfs-DWI provides better subjective and objective image quality than standard SPAIR-DWI, resulting in a more accurate estimation of benign and malignant lesion size.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.acra.2023.01.014 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!