AI Article Synopsis

  • The study aimed to assess how accurately glaucoma worsening can be detected over two years using different frequencies of Optical Coherence Tomography (OCT) scans and visual field (VF) tests in a large population of adults with glaucoma or at risk for it.
  • Researchers analyzed data from over 12,000 eyes using OCT and 20,000 eyes using VF from patients monitored between 2013 and 2021, focusing on rates of change in retinal nerve fiber layer thickness and mean deviation.
  • Findings revealed that the accuracy of detecting moderate and rapid glaucoma worsening was under 50%, but using both OCT and VF together significantly improved detection accuracy by over 17 percentage points compared to using either method alone

Article Abstract

Purpose: To compare the accuracy of detecting moderate and rapid rates of glaucoma worsening over a 2-year period with different numbers of OCT scans and visual field (VF) tests in a large sample of glaucoma and glaucoma suspect eyes.

Design: Descriptive and simulation study.

Participants: The OCT sample comprised 12 150 eyes from 7392 adults with glaucoma or glaucoma suspect status followed up at the Wilmer Eye Institute from 2013 through 2021. The VF sample comprised 20 583 eyes from 10 958 adults from the same database. All eyes had undergone at least 5 measurements over follow-up from the Zeiss Cirrus OCT or Humphrey Field Analyzer.

Methods: Within-eye rates of change in retinal nerve fiber layer (RNFL) thickness and mean deviation (MD) were measured using linear regression. For each measured rate, simulated measurements of RNFL thickness and MD were generated using the distributions of residuals. Simulated rates of change for different numbers of OCT scans and VF tests over a 2-year period were used to estimate the accuracy of detecting moderate (75th percentile) and rapid (90th percentile) worsening for OCT and VF. Accuracy was defined as the percentage of simulated eyes in which the true rate of worsening (the rate without measurement error) was at or less than a criterion rate (e.g., 75th or 90th percentile).

Main Outcome Measures: The accuracy of diagnosing moderate and rapid rates of glaucoma worsening for different numbers of OCT scans and VF tests over a 2-year period.

Results: Accuracy was less than 50% for both OCT and VF when diagnosing worsening after a 2-year period. OCT accuracy was 5 to 10 percentage points higher than VF accuracy at detecting moderate worsening and 10 to 15 percentage points higher for rapid worsening. Accuracy increased by more than 17 percentage points when using both OCT and VF to detect worsening, that is, when relying on either OCT or VF to be accurate.

Conclusions: More frequent OCT scans and VF tests are needed to improve the accuracy of diagnosing glaucoma worsening. Accuracy greatly increases when relying on both OCT and VF to detect worsening.

Financial Disclosure(s): The author(s) have no proprietary or commercial interest in any materials discussed in this article.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10200740PMC
http://dx.doi.org/10.1016/j.ophtha.2023.01.021DOI Listing

Publication Analysis

Top Keywords

oct scans
20
glaucoma worsening
16
oct
13
accuracy detecting
12
detecting moderate
12
2-year period
12
numbers oct
12
scans tests
12
percentage points
12
worsening
10

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!