Purpose: To examine the rotational stability of a new toric intraocular lens (IOL), the TECNIS Eyhance Toric II, over a course of 3 months.
Setting: Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria.
Design: Prospective unmasked single-center study.
Methods: 50 eyes of 50 patients with cataract and regular corneal astigmatism ≥0.75 diopters (D) were included. The TECNIS Eyhance Toric II IOL was implanted in 1 eye of each study patient. Images of the alignment axis of the IOL were taken intraoperatively, as well as at 1 hour, 1 week, and 3 months postoperatively. For the measurement of toric IOL rotation, images were superimposed on the basis of ocular landmarks. Uncorrected and corrected distance (4 m) and intermediate (66 cm) visual acuities were assessed at the 3-month visit.
Results: There was no significant difference in the rotational position between the intraoperative and 3 month timepoints, with a mean rotation of 1.34 ± 1.46 degrees, in 27 examined eyes. ( P = 0.313). No patients had to undergo repositioning of the toric IOL. A significant reduction of refractive cylinder occurred from preoperatively 1.8 ± 1.1 to 0.40 ± 0.42 D at the 3-month visit ( P = .001; n = 43). The corrected distance visual acuity improved significantly from 0.28 ± 0.16 logMAR preoperatively to -0.01 ± 0.13 logMAR at 3 months postoperatively ( P = .001; n = 43).
Conclusions: The TECNIS Eyhance Toric II showed a good visual performance with no significant rotation over a course of 3 months and, therefore, an excellent rotational stability. The intraocular lens showed a good safety profile with no adverse events.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/j.jcrs.0000000000001158 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!