Purpose: To evaluate pseudo-continuous arterial spin labeling (pCASL) and velocity-selective arterial spin labeling (VSASL) for quantification of spinal cord blood flow (SCBF) in the rat thoracolumbar spinal cord.
Methods: Labeling efficiency (LE) was compared between pCASL and three VSASL variants in simulations and both phantom and in vivo experiments at 9.4 T. For pCASL, the effects of label plane position and shimming were systematically evaluated. For VSASL, the effects of composite pulses and phase cycling were evaluated to reduce artifacts. Additionally, vessel suppression, respiratory, and cardiac gating were evaluated to reduce motion artifacts. pCASL and VSASL maps of spinal cord blood flow were acquired with the optimized protocols.
Results: LE of the descending aorta was larger in pCASL compared to VSASL variants. In pCASL, LE off-isocenter was improved by local shimming positioned at the label plane and the anatomical level of labeling for the thoracic cord was only viable at the level of the T10 vertebra. Cardiac gating was essential to reduce motion artifacts. Both pCASL and VSASL successfully demonstrated comparable SCBF values in the thoracolumbar cord.
Conclusion: pCASL demonstrated high and consistent LE in the thoracic aorta, and VSASL was also feasible, but with reduced efficiency. A combination of cardiac gating and recording of actual post-label delays was important for accurate SCBF quantification. These results highlight the challenges and solutions to achieve sufficient ASL labeling and contrast at high field in organs prone to motion.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10050093 | PMC |
http://dx.doi.org/10.1002/mrm.29603 | DOI Listing |
Z Med Phys
July 2024
Divison of Radiology, German Cancer Research Center, Heidelberg, Germany; Division of Neuroradiology, Geneva University Hospitals, Geneva, Switzerland. Electronic address:
Purpose: To apply velocity selective arterial spin labeling (VSASL) combined with a navigator-based (NAV) prospective motion compensation method for a free-breathing liver perfusion measurement without contrast agent.
Methods: Sinc-modulated Velocity Selective Inversion (sinc-VSI) pulses were applied as labeling and control pulses. In order to account for respiratory motion, a navigator was employed in the form of a single gradient-echo projection readout, located at the diaphragm along the inferior-superior direction.
Magn Reson Med
November 2024
The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University, Baltimore, Maryland, USA.
Purpose: Velocity selective arterial spin labeling (VSASL) quantification assumes that the labeled bolus continuously moves into the imaging voxel during the post-labeling delay (PLD). Faster blood flow could lead to a bolus duration shorter than the applied PLD of VSASL and cause underestimation of cerebral blood flow (CBF). This study aims to evaluate the performance of velocity-selective inversion (VSI) prepared arterial spin labeling (ASL) with different PLDs and pseudo-continuous ASL (PCASL) for quantification of hypercapnia-induced cerebrovascular reactivity (CVR), using phase-contrast (PC) MRI as a global reference.
View Article and Find Full Text PDFMagn Reson Med
June 2024
Department of Bioengineering, University of California Riverside, Riverside, California, USA.
Purpose: Background suppression (BS) is recommended in arterial spin labeling (ASL) for improved SNR but is difficult to optimize in existing velocity-selective ASL (VSASL) methods. Dual-module VSASL (dm-VSASL) enables delay-insensitive, robust, and SNR-efficient perfusion imaging, while allowing efficient BS, but its optimization has yet to be thoroughly investigated.
Methods: The inversion effects of the velocity-selective labeling pulses, such as velocity-selective inversion (VSI), can be used for BS, and were modeled for optimizing BS in dm-VSASL.
Magn Reson Med
June 2023
Department of Neurosurgery, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.
Purpose: To evaluate pseudo-continuous arterial spin labeling (pCASL) and velocity-selective arterial spin labeling (VSASL) for quantification of spinal cord blood flow (SCBF) in the rat thoracolumbar spinal cord.
Methods: Labeling efficiency (LE) was compared between pCASL and three VSASL variants in simulations and both phantom and in vivo experiments at 9.4 T.
Eur Radiol
May 2022
Department of Radiology, Zhujiang Hospital, Southern Medical University, 253 Gongye Middle Avenue, Haizhu District, Guangzhou, 510282, Guangdong, China.
Objectives: To evaluate the performance of velocity-selective (VS) ASL among patients with untreated gliomas by comparing with both pseudo-continuous (PC) ASL and dynamic susceptibility contrast-enhanced perfusion-weighted imaging (DSC-PWI).
Methods: Forty-four consecutive patients with newly diagnosed glioma who underwent preoperative perfusion MRI including VSASL, PCASL, and DSC-PWI between 2017 and 2019 were retrospectively evaluated. Visual inspection was performed to evaluate the tumor signal intensity relative to gray matter based on 1-5 score criteria and weighted kappa was used to evaluate the pair-wise concordance between VSASL or PCASL and DSC-PWI.