Introduction: The recent introduction of the European Medical Device Regulation poses stricter legislation for manufacturers developing medical devices in the EU. Many devices have been placed into a higher risk category, thus requiring more data before market approval, and a much larger focus has been placed on safety. For implantable and Class III devices, the highest risk class, clinical evidence is a necessity. However, the requirements of clinical study design and developmental outcomes are only described in general terms due to the diversity of devices.

Methods: A structured approach to determining the requirements for the clinical development of high-risk medical devices is introduced, utilizing the question-based development framework, which is already used for pharmaceutical drug development. An example of a novel implantable device for haemodialysis demonstrates how to set up a relevant target product profile defining the device requirements and criteria. The framework can be used in the medical device design phase to define specific questions to be answered during the ensuing clinical development, based upon five general questions, specified by the question-based framework.

Results: The result is a clear and evaluable overview of requirements and methodologies to verify and track these requirements in the clinical development phase. Development organizations will be guided to the optimal route, also to abandon projects destined for failure early on to minimize development risks.

Conclusion: The framework could facilitate communication with funding agencies, regulators and clinicians, while highlighting remaining 'known unknowns' that require answering in the post-market phase after sufficient benefit is established relative to the risks.

Download full-text PDF

Source
http://dx.doi.org/10.1111/bcp.15685DOI Listing

Publication Analysis

Top Keywords

medical devices
12
requirements clinical
12
clinical development
12
question-based development
8
development high-risk
8
high-risk medical
8
medical device
8
development
7
medical
5
devices
5

Similar Publications

Electromyography (EMG) is increasingly used in stroke assessment research, with studies showing that EMG co-contraction (EMG-CC) of upper limb muscles can differentiate stroke patients from healthy individuals and correlates with clinical scales assessing motor function. This suggests that EMG-CC has potential for both assessing motor impairments and monitoring recovery in stroke patients. However, systematic reviews on EMG-CC's effectiveness in stroke assessment are lacking.

View Article and Find Full Text PDF

Randomized clinical trials and clinical practice guidelines recommend the use of self-measured blood pressure (SMBP) to help improve the treatment of patients with hypertension. Many clinicians use SMBP in their practices, but there is significant variability in how SMBP is implemented in their day-to-day practice. This quality improvement study details the pragmatic and real-world approach clinicians and administrators used at 3 sites of the IHA Medical Group, a part of Trinity Healthcare, to implement the American Medical Association (AMA) 7-Step SMBP framework as part of the larger AMA hypertension quality improvement program AMA MAP BP.

View Article and Find Full Text PDF

Patient-reported outcome measures (PROMs) are vital tools in cardiovascular disease (CVD) research and care, providing insights that complement traditional clinical outcomes like mortality and morbidity. PROMs capture patient experiences with CVD, such as quality of life, functional capacity, and emotional well-being, allowing clinicians to assess how interventions impact daily life. PROMs are integral to cardiovascular investigations as well as management, especially in chronic conditions and rehabilitation, where they inform on the impact of personalized care plans by tracking symptom progression and patient adherence.

View Article and Find Full Text PDF

The impact of anions on electrooxidation of perfluoroalkyl acids by porous Magnéli phase titanium suboxide anodes.

PLoS One

January 2025

Department of Crop and Soil Sciences, College of Agricultural and Environmental Sciences, University of Georgia, Griffin, Georgia, United States of America.

Previous studies have indicated the great performance of electrooxidation (EO) to mineralize per- and polyfluoroalkyl substances (PFASs) in water, but different anions presented in wastewater may affect the implementation of EO treatment in field applications. This study invetigated EO treatment of perfluorooctane sulfonate (PFOS) and perfluorooctanoic acid (PFOA), two representative perfluoroalkyl acids (PFAAs), using porous Magnéli phase titanium suboxide anodes in electrolyte solutions with different anions present, including NO3-, SO42-, CO32- and PO43-. The experiment results indicate that CO32- enhanced PFAS degradation, while NO3- suppressed the degradation reactions with its concentration higher than 10 mM.

View Article and Find Full Text PDF

Disposable filtering face piece respirators (FFRs) are not approved for reuse as standard of care. However, lessons learnt from the SARS-CoV-2 pandemic, FFRs decontamination and reuse may be needed as crisis capacity strategy to ensure availability in medical facilities. We studied a decontamination methodology based on atmospheric pressure plasma technology, which allows for rapid, contact-free decontamination without utilisation of harmful chemicals, and suitable to access small pores and microscopic filters openings.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!