Comparison of ionised calcium measured using a portable analyser to a reference method in healthy dogs.

J Small Anim Pract

College of Public Health, Medical and Veterinary Sciences, James Cook University, Townsville, Australia.

Published: May 2023

Objectives: To compare the ionised calcium measured on a portable analyser (iSTAT, Abbott) to a reference method.

Materials And Methods: Blood samples from 39 apparently healthy dogs were analysed in duplicate using a portable analyser and a reference method (Radiometer ABL800 FLEX). Bland-Altman plots and Passing-Bablok regression were used to assess constant and proportional bias between the two instruments. A within-assay percentage coefficient of variation and total error (TE) was calculated for both analysers. The reference interval was calculated for the portable analyser using the robust method with confidence interval bootstrapping.

Results: The Bland-Altman plot showed a -0.036 mmol/L difference between the two instruments (95% confidence limit -0.08 to 0.01 mmol/L; limits of agreement -0.07 to 0.006 mmol/L). Neither the Bland-Altman plot nor the Passing-Bablock regression (slope -0.03; 95% confidence interval -0.08 to 0.19 and intercept 1; 95% confidence interval 0.83 to 1.2) showed significant proportional bias. The coefficient of variation for the portable analyser was 1.08%, compared to 0.78% for the reference method with a total error of 3.5% for the portable analyser. The estimated population-based reference interval for ionised calcium using the portable analyser is 1.23 to 1.42 mmol/L.

Clinical Significance: For the healthy dogs in this study, compared to the reference method, the portable analyser showed no significant bias for measurement of ionised calcium. Further studies including hyper and hypocalcaemic dogs are required to determine clinical impact of the use of this analyser.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jsap.13588DOI Listing

Publication Analysis

Top Keywords

portable analyser
32
ionised calcium
16
reference method
16
healthy dogs
12
confidence interval
12
95% confidence
12
analyser
9
calcium measured
8
portable
8
measured portable
8

Similar Publications

The calcium requirements of dairy cows increase dramatically soon after calving, and many cows have subclinical hypocalcemia, adversely affecting health and performance. Traditional laboratory tests for calcium are complex and not easily adapted to rapid point-of-care applications. The objectives were to evaluate a portable iCa testing device, Horiba LAQUAtwin Ca-11C, for measuring ionized calcium (iCa) in the whole blood of dairy cows and to investigate the iCa-to-total-calcium (tCa) ratio in blood collected from dairy cows within 9 days after calving.

View Article and Find Full Text PDF

High through-put groundwater arsenic speciation analysis using an automated flow analyzer.

J Environ Sci (China)

July 2025

State Key Laboratory of Marine Environmental Science, College of the Environment and Ecology, Xiamen University, Xiamen 361102, China; National Observation and Research Station for the Taiwan Strait Marine Ecosystem, Xiamen University, Zhangzhou 363000, China. Electronic address:

The occurrence of geogenic arsenic (As) in groundwater is a global public health concern. However, there remain large gaps in groundwater As data, making it difficult to identify non-compliant domestic wells, partly due to lack of low-cost methods capable of rapid As analysis. Therefore, the development of high through-put and reliable on-site determination methods for inorganic As is essential.

View Article and Find Full Text PDF

Purpose: The purpose of this systematic review was to consolidate and summarize available data comparing virtual reality perimetry (VRP) with standard automated perimetry (SAP) in adults with glaucoma. Understanding the utility and diagnostic performance of emerging VRP technology may expand access to visual field testing but requires evidence-based validation.

Methods: A systematic literature search was conducted in 3 databases (PubMed Central, Embase, and Cochrane Central Register of Controlled Trials) from the date of inception to 10/09/2024.

View Article and Find Full Text PDF

Introduction: Hemoglobin (Hb) estimation in blood donors is conducted using capillary samples on portable hemoglobinometers, representing measurement methods in practice. The reference standard is conducted using a venous sample on a hematology analyzer, representing the mentor measurement method or the true value. The correction involves the calculation of the secondary adjustment factor (SAF) to mitigate the difference between the two values.

View Article and Find Full Text PDF

An FPGA-Based SiNW-FET Biosensing System for Real-Time Viral Detection: Hardware Amplification and 1D CNN for Adaptive Noise Reduction.

Sensors (Basel)

January 2025

Department of Computer Science, Faculty of Sciences and Humanities Sciences, Majmaah University, Al Majmaah 11952, Saudi Arabia.

Impedance-based biosensing has emerged as a critical technology for high-sensitivity biomolecular detection, yet traditional approaches often rely on bulky, costly impedance analyzers, limiting their portability and usability in point-of-care applications. Addressing these limitations, this paper proposes an advanced biosensing system integrating a Silicon Nanowire Field-Effect Transistor (SiNW-FET) biosensor with a high-gain amplification circuit and a 1D Convolutional Neural Network (CNN) implemented on FPGA hardware. This attempt combines SiNW-FET biosensing technology with FPGA-implemented deep learning noise reduction, creating a compact system capable of real-time viral detection with minimal computational latency.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!