Ultra-low-dose CT is feasible for torsion measurement of the lower limb in patients with metal implants.

Br J Radiol

Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Eberhard Karls University Tübingen, Hoppe-Seyler-Str, Tübingen, Germany.

Published: March 2023

Objectives: Patients who need torsion measurement of the lower limb often have metal implants hindering . MRI. A new ultra-low-dose (ULD-)CT protocol might be feasible for torsion measurement at cost of relatively low radiation exposure.

Methods: We retrospectively included all patients with clinically indicated torsion measurement in the period July 2019 to June 2021 and metal implants in the scanning field. The ULD-CT protocol comprised automated tube current time product and automated tube voltage with reference settings of 100kV/20mAs (hip), 80kV/20mAs (knee) and 80kV/10mAs (ankle). Femoral neck anteversion, tibial, intra-articular knee and overall leg torsion measurements were performed by two radiologists independently. Diagnostic confidence regarding the delineation of the relevant cortical bone was rated on a 5-point Likert scale (1 = non-diagnostic, 5 = excellent).

Results: 102 consecutive patients could be included (BMI 27.38 ± 5.85) with 154 metal implants. Median total dose length product of the ULD-CT-torsion measurement was 16.5mGycm [11-39]. Both readers showed high agreement with a maximum torsional difference of 4.1°. Diagnostic confidence was rated best (5/5) in 92.2% (reader 1) and 93.1% (reader 2) with a worst rating of 3/5.

Conclusion: The new ULD-CT protocol is feasible for torsion measurement of the lower limb - even in patients with metal implants.

Advances In Knowledge: Metal implants are not an obstacle for ULD-CT torsion measurements of the lower limb.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10078873PMC
http://dx.doi.org/10.1259/bjr.20220495DOI Listing

Publication Analysis

Top Keywords

torsion measurement
20
metal implants
20
lower limb
16
feasible torsion
12
measurement lower
12
uld-ct protocol
12
limb patients
8
patients metal
8
protocol feasible
8
automated tube
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!