Background: Mitral valve transcatheter edge-to-edge repair is safe and effective in treating degenerative mitral regurgitation (DMR) patients at prohibitive surgical risk, but outcomes in complex mitral valve anatomy patients vary.

Objectives: The PASCAL IID registry assessed safety, echocardiographic, and clinical outcomes with the PASCAL system in prohibitive risk patients with significant symptomatic DMR and complex mitral valve anatomy.

Methods: Patients in the prospective, multicenter, single-arm registry had 3+ or 4+ DMR, were at prohibitive surgical risk, presented with complex anatomic features based on the MitraClip instructions for use, and were deemed suitable for the PASCAL system by a central screening committee. Enrolled patients were treated with the PASCAL system. Safety, effectiveness, and functional and quality-of-life outcomes were assessed. Study oversight also included an echocardiographic core laboratory and clinical events committee.

Results: The study enrolled 98 patients (37.2% ≥2 independent significant jets, 15.0% severe bileaflet/multi scallop prolapse, 13.3% mitral valve orifice area <4.0 cm, and 10.6% large flail gap and/or large flail width). The implant success rate was 92.9%. The 30-day composite major adverse event rate was 11.2%. At 6 months, 92.4% patients achieved MR ≤2+ and 56.1% achieved MR ≤1+ (P < 0.001 vs baseline). The Kaplan-Meier estimates for survival, freedom from major adverse events, and heart failure hospitalization at 6 months were 93.7%, 85.6%, and 92.6%, respectively. Patients experienced significant symptomatic improvement compared with baseline (P < 0.001).

Conclusions: The outcomes of the PASCAL IID registry establish the PASCAL system as a useful therapy for prohibitive surgical risk DMR patients with complex mitral valve anatomy. (PASCAL IID Registry within the Edwards PASCAL TrAnScatheter Valve RePair System Pivotal Clinical Trial [CLASP IID] NCT03706833).

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jacc.2022.11.034DOI Listing

Publication Analysis

Top Keywords

mitral valve
16
pascal system
12
transcatheter edge-to-edge
8
edge-to-edge repair
8
degenerative mitral
8
mitral regurgitation
8
prohibitive surgical
8
surgical risk
8
complex mitral
8
enrolled patients
8

Similar Publications

Outcomes and Impact of Device Iterations in Mitral Valve Transcatheter Edge-to-Edge Repair: The REPAIR Study.

JACC Cardiovasc Interv

November 2024

Department of Cardiology, Heart Center, Faculty of Medicine, University of Cologne, Cologne, Germany. Electronic address:

Background: The PASCAL P10 system for mitral valve transcatheter edge-to-edge repair has undergone iterations, including introduction of the narrower Ace implant and the Precision delivery system.

Objectives: The study sought to evaluate outcomes and the impact of PASCAL mitral valve transcatheter edge-to-edge repair device iterations.

Methods: The REPAIR (REgistry of PAscal for mltral Regurgitation) study is an investigator-initiated, multicenter registry including consecutive patients with mitral regurgitation (MR) treated from 2019 to 2024.

View Article and Find Full Text PDF

Cardiac MRI Evaluation of Determinants and Prognostic Implications of Right Ventricular Dysfunction in Aortic Regurgitation.

Radiol Cardiothorac Imaging

February 2025

From the Department of Cardiology, Houston Methodist DeBakey Heart & Vascular Center, 6550 Fannin St, Smith Tower, Ste 1801, Houston, TX 77030 (M.M., P.B., V.C., M.S., M.R., S.F.N., W.A.Z., D.J.S.); and Department of Pathology and Genomic Medicine, Houston Methodist Hospital Research Institute, Houston, Tex (D.T.N., E.A.G.).

Purpose To investigate the determinants and effect of right ventricular (RV) dysfunction in aortic regurgitation (AR) using cardiac MRI. Materials and Methods This study included patients with moderate or severe AR who were enrolled in the DEBAKEY-CMR registry between January 2009 and June 2020. Patients with previous valve intervention, cardiomyopathy deemed unrelated to AR, severe aortic stenosis, and other confounders were excluded.

View Article and Find Full Text PDF

Objectives: Surgical options for mitral valve repair in dogs are unstandardized and influenced by various factors. This study describes a four-point surgical technique (MI-4) to reduce mitral regurgitation and provides data from a study of dogs with 12 months of follow up.

Methods: Twenty-five dogs with stages C or D mitral insufficiency were treated by one of two surgeons using the MI-4 procedure at Ueno no Mori Animal Hospital between October 2021 and May 2023.

View Article and Find Full Text PDF

True- and pseudo-mitral annular disjunction in patients undergoing cardiovascular magnetic resonance.

J Cardiovasc Magn Reson

December 2024

IRCCS Humanitas Research Hospital, Via Alessandro Manzoni, 56, 20089 Rozzano, Milano, Italy; Department of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini, 4, 20090 Pieve Emanuele, Milano, Italy. Electronic address:

Background: Mitral annular disjunction (MAD) is a controversial entity. Recently, a distinction between pseudo-MAD, present in systole and secondary to juxtaposition of the billowing posterior leaflet on the left atrial wall, and true-MAD, where the insertion of the posterior leaflet is displaced on the atrial wall both in diastole or in systole, has been proposed. We investigated the prevalence of pseudo-MAD and true-MAD.