  • - Researchers developed a non-invasive brain pulse monitor that uses red light to detect blood flow signals from the brain's surface, offering a safer and cheaper alternative to traditional invasive intracranial pressure (ICP) monitoring.
  • - A study involving 12 critically ill patients showed that the brain pulse monitor's signals closely resembled those of invasive ICP measurements, with both methods indicating increased amplitude in response to higher ICP levels.
  • - An automated algorithm confirmed a strong correlation (R=0.66, P < 0.001) between the brain pulse waveform patterns and ICP levels, suggesting that this new monitoring technique could effectively measure ICP without surgery.

Article Abstract

Background: Intracranial pressure (ICP) monitoring requires placing a hole in the skull through which an invasive pressure monitor is inserted into the brain. This approach has risks for the patient and is expensive. We have developed a non-invasive brain pulse monitor that uses red light to detect a photoplethysmographic (PPG) signal arising from the blood vessels on the brain's cortical surface. The brain PPG and the invasive ICP waveform share morphological features which may allow measurement of the intracranial pressure.

Methods: We enrolled critically ill patients with an acute brain injury with invasive ICP monitoring to assess the new monitor. A total of 24 simultaneous invasive ICP and brain pulse monitor PPG measurements were undertaken in 12 patients over a range of ICP levels.

Results: The waveform morphologies were similar for the invasive ICP and brain pulse monitor PPG approach. Both methods demonstrated a progressive increase in the amplitude of P2 relative to P1 with increasing ICP levels. An automated algorithm was developed to assess the PPG morphological features in relation to the ICP level. A correlation was demonstrated between the brain pulse waveform morphology and ICP levels, R=0.66, P < 0.001.

Conclusion: The brain pulse monitor's PPG waveform demonstrated morphological features were similar to the invasive ICP waveform over a range of ICP levels, these features may provide a method to measure ICP levels.

Trial Registration: ACTRN12620000828921.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9883992PMC
http://dx.doi.org/10.2147/MDER.S398193DOI Listing

Publication Analysis

Top Keywords

brain pulse
24
invasive icp
20
pulse monitor
16
icp
12
morphological features
12
icp levels
12
brain
10
non-invasive brain
8
acute brain
8
brain injury
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!