Background: Hip and knee flexion joint motions are frequently examined in clinical practice using camera based motion capture (CBMC) systems; however, these systems require elaborate setups and dedicated space. Inertial measurement unit (IMU) based systems avoid these disadvantages but require validation before widespread adoption. Moreover, it is important for clinical practice to determine the stability of these systems for prolonged evaluation periods. The purpose of this study was to assess the validity of a three-sensor inertial measurement unit system for calculating hip and knee flexion angles during gait by comparing with a gold standard CBMC system. Validity was also examined before and after a treadmill walking session.

Methods: Twenty healthy participants were tested. Twenty seconds of gait at preferred walking speed were analyzed before and after thirty-two minutes of treadmill walking using previously validated CBMC methods and with a custom IMU model. Measurement validity for the IMU system was evaluated using Bland & Altman 95 percent limits of agreement, linear regression, mean absolute error and root mean square error. The effects of a measurement zeroing calibration strategy were also investigated.

Results: Strong measurement agreement was observed for both hip and knee flexion angles, although overall agreement for the hip exceeded that for the knee. Linear regressions between the datasets for each participant illustrated strong (> 0.94) relationships between IMU and CBMC measurements. More significant changes between timepoints were observed for the knee than for the hip. Error values were generally reduced when zeroing calibration was implemented.

Conclusion: The IMU system presented in this study is a convenient and accessible technique to measure joint angles. The protocol described in the current study can be easily applied in the clinical setting for evaluation of clinical populations. Additional development work on sensor placement and calibration methods may further increase the accuracy of such methods. Clinical translation statement: The IMU system presented in this study is a convenient and accessible technique to measure joint angles. Additional developmentwork on sensor placement and calibration methods may further increase the accuracy of such methods.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9870263PMC
http://dx.doi.org/10.1109/JTEHM.2022.3226153DOI Listing

Publication Analysis

Top Keywords

inertial measurement
12
measurement unit
12
hip knee
12
knee flexion
12
imu system
12
knee hip
8
clinical practice
8
flexion angles
8
treadmill walking
8
zeroing calibration
8

Similar Publications

Background: Global technological advancements have shifted equine lameness diagnostics from purely subjective assessment toward more objective, quantitative methods. Traditional gait evaluations are increasingly being supplemented by innovative technologies to enhance diagnostic accuracy.

Aim: This study aimed to determine if traditional gait assessment could be effectively complemented by quantitative measurements using an affordable, self-constructed device, the Lameness Detector 0.

View Article and Find Full Text PDF

Purpose: Motion capture technology is quickly evolving providing researchers, clinicians, and coaches with more access to biomechanics data. Markerless motion capture and inertial measurement units (IMUs) are continually developing biomechanics tools that need validation for dynamic movements before widespread use in applied settings. This study evaluated the validity of a markerless motion capture, IMU, and red, green, blue, and depth (RGBD) camera system as compared to marker-based motion capture during countermovement jumps, overhead squats, lunges, and runs with cuts.

View Article and Find Full Text PDF

Computer model for gait assessments in Parkinson's patients using a fuzzy inference model and inertial sensors.

Artif Intell Med

December 2024

Escuela Nacional de Medicina y Homeopatía, Instituto Politécnico Nacional, Guillermo Massieu 239, 07320 Mexico City, Mexico.

Patients with Parkinson's disease (PD) in the moderate and severe stages can present several walk alterations. They can show slow movements and difficulty initiating, varying, or interrupting their gait; freezing; short steps; speed changes; shuffling; little arm swing; and festinating gait. The Movement Disorder Society-Unified Parkinson's Disease Rating Scale (MDS-UPDRS) has a good reputation for uniformly evaluating motor and non-motor aspects of PD.

View Article and Find Full Text PDF

Assessment of varus thrust using inertial measurement units.

Clin Biomech (Bristol)

December 2024

Department of Kinesiology and Applied Physiology, University of Delaware, Newark, DE, USA. Electronic address:

Background: Varus thrust is common in those with knee osteoarthritis. Varus thrust is traditionally identified with visual analysis or motion capture, methods that are either dichotomous or limited to the laboratory setting. Inertial measurement unit data has been found to correlate with motion capture measures of varus thrust in those with severe knee osteoarthritis, allowing for a quantitative and accessible way of measuring varus thrust.

View Article and Find Full Text PDF

GPS/VIO integrated navigation system based on factor graph and fuzzy logic.

Sci Rep

December 2024

Department of Electrical Engineering, Iran University of Science and Technology, Tehran, 16846-13114, Iran.

In today's technologically advanced landscape, precision in navigation and positioning holds paramount importance across various applications, from robotics to autonomous vehicles. A common predicament in location-based systems is the reliance on Global Positioning System (GPS) signals, which may exhibit diminished accuracy and reliability under certain conditions. Moreover, when integrated with the Inertial Navigation System (INS), the GPS/INS system could not provide a long-term solution for outage problems due to its accumulated errors.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!