Purpose: To evaluate the prediction error (PE) variance and absolute median PE of different intraocular lens (IOL) calculation formulas including last-generation formulas such as Barrett True-K with K, Okulix and total keratometry (TK)-based calculations with Haigis, and Barrett True-K in a simulation model in post-small-incision lenticule extraction (SMILE) eyes.
Settings: Department of Ophthalmology, University Hospital Marburg, Marburg, Germany.
Design: Prospective study.
Methods: Preoperative measurements included IOL power calculation before and after SMILE surgery. The target refraction was set to be the lowest myopic refractive error in pre-SMILE eyes. The IOL power targeting at the lowest myopic refractive error in pre-SMILE eyes was selected for the post-SMILE IOL calculation of the same eye. The difference between the predicted refraction of pre- and post-SMILE eyes with the same IOL power was defined as IOL difference. The refractive change induced by SMILE was defined as the difference between preoperative and postoperative manifest refraction.
Results: 98 eyes from 49 patients underwent bilateral myopic SMILE. The PE variance of Okulix was not significantly different compared with Barrett True-K with TK ( P = .471). The SDs of the mean PEs were ±0.413 D (Haigis-TK), ±0.453 D (Okulix), ±0.471 D (Barrett True-K with TK), ±0.556 D (Haigis-L), and ±0.576 D (Barrett True-K with K). The mean absolute PE was 0.340 D, 0.353 D, 0.404 D, 0.511 D, and 0.715 D for Haigis-TK, Okulix, Barrett True-K with TK, Barrett True-K with K, and Haigis-L, respectively. The highest percentage of eyes within ±0.50 D was achieved by Okulix, followed by Haigis-TK, Barrett True-K with TK, Barrett True-K with K, and Haigis-L.
Conclusions: Results suggest that Haigis in combination with TK, Okulix, and Barrett True-K with and without TK offer good options for accurate IOL power calculation after SMILE. Haigis-L showed a tendency for myopic shift in eyes after previous SMILE.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/j.jcrs.0000000000001139 | DOI Listing |
Int Ophthalmol
January 2025
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology and Visual Sciences Key Laboratory, No.1, Dong Jiao Min Xiang, Dong Cheng District, Beijing, 100730, China.
Purpose: To compare the accuracy of intraocular lens (IOL) power calculation formulas in cataract patients with keratoconus (KC).
Methods: This study followed the Preferred Reporting Items for Systematic Reviews and Meta-analysis statementand and was registered on PROSPERO (CRD42024568997). Pubmed, Web of Science, Cochrane Library, and EMBASE were searched for retrospective and prospective clinical studies published until October 2024.
Clin Ophthalmol
December 2024
Alcon Vision LLC, Fort Worth, TX, USA.
Purpose: To compare the refractive prediction accuracy of the Optiwave Refractive Analysis (ORA) SYSTEM with the Barrett True-K (BTK) formula in calculating intraocular lens (IOL) power in eyes that underwent cataract surgery after previous myopic photorefractive keratectomy (PRK) or laser-assisted in situ keratomileusis (LASIK).
Methods: This retrospective study evaluated patients aged ≥22 years with prior myopic PRK or LASIK who underwent unilateral or bilateral cataract removal and monofocal IOL implantation using the ORA SYSTEM at 177 sites in the United States. Two datasets were analyzed: All Eyes (ie, all eligible eyes) and First Surgery Eyes (ie, each patient's first implanted eye).
Clin Exp Ophthalmol
December 2024
Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Background: Accurate intraocular lens (IOL) calculation in eyes with keratoconus (KCN) poses significant challenges. While various formulas, including KCN-specific ones, have been investigated, the optimal calculation method remains inconclusive.
Methods: The study was pre-registered in PROSPERO (CRD42023483119).
Int Ophthalmol
November 2024
Anterior Segment Surgery Department, Asociación Para Evitar La Ceguera en México I.A.P., Vicente García Torres 46, Barrio San Lucas, CP 04030, Coyoacán, Mexico City, Mexico.
Purpose: To describe the accuracy of monofocal intraocular lens power calculation in patients with keratoconus using total keratometry (TK) and standard keratometry (K) with conventional and keratoconus-modified formulas.
Setting: Asociación Para Evitar la Ceguera en México, Mexico City, Mexico.
Design: Observational, retrospective, non-randomized, comparative study.
Sci Rep
November 2024
Department of Ophthalmology, Keio University School of Medicine, Tokyo, Japan.
This retrospective study compared postoperative prediction errors of recent formulas using standard- or total keratometry (K or TK) for intraocular lens (IOL) power calculation in post-myopic LASIK patients. It included 56 eyes of 56 patients who underwent uncomplicated cataract surgery, with at least 1-month follow-up at Keio University Hospital in Tokyo or Hayashi Eye Hospital in Yokohama, Japan. Prediction errors, absolute errors, and percentage of eyes with prediction errors within ± 0.