Aim: The aim of this study was to measure the dose to planning target and organ at risk (OAR) using Alderson Rando phantom for various treatment techniques in left breast radiotherapy and to estimate the secondary cancer incidence.
Materials And Methods: Eleven different combinations of plans containing four techniques (three dimensional conformal radiotherapy, intensity-modulated radiation therapy [IMRT], volumetric modulated arc therapy [VMAT], and combination of 3DCRT and VMAT plans (HYBRID)) were created with 6 MV FF and 6 MV FFF (flattening filter and flattening filter-free) photon energies in phantom. Planned target volume and OAR doses in 23 different locations were measured using optically stimulated luminescence dosimeter (OSLD) and EBT3 films. Assuming the age of exposure as 30 years, lifetime attributable risk (LAR) was estimated based on excess absolute risk (EAR) models outlined in the Biological Effects of Ionizing Radiation VII report.
Results: Film showed maximum deviations of 6.15% with IMRT_C_FF plan when compared with treatment planning system (TPS). The maximum percentage difference of 1.7% was found with OSLD measurement when compared with TPS for VMAT_T_FFF plan. EAR estimation was done for all the OARs including target. The LARs for left lung, right lung, and right breast were evaluated. The maximum LAR values of 2.92 ± 0.14 were found for left lung with VMAT_C_FFF plans.
Conclusion: This study shows that both OSLD and EBT3 films are suitable for dose measurements using Rando phantom. OSLD shows superior results when compared with films, especially with relatively larger distances. Maximum LAR values were found with VMAT_C_FFF plans. Considering the secondary cancer risk associated with the patients treated in the younger age group, it is suggested that dose estimation should be a part of treatment quality audit whenever possible.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9847007 | PMC |
http://dx.doi.org/10.4103/jmp.jmp_36_22 | DOI Listing |
Sci Rep
March 2025
Medical Physics and Informatics Laboratory of Electronics Engineering, National Kaohsiung University of Sciences and Technology, Kaohsiung, 80778, Taiwan, ROC.
In radiation therapy, precise dose distribution is essential for minimizing damage to normal tissues. Gafchromic EBT3 film is widely used to assure the quality of two-dimensional dosimetry but requires frequent recalibrations due to changes in sensitivity over time. This study presents a new calibration method using a Keras-based generalized additive neural network (GANN) to address film aging.
View Article and Find Full Text PDFFront Oncol
December 2024
Institute of Radiation Medicine (IRM), Helmholtz Zentrum München GmbH, German Research Center for Environmental Health, Neuherberg, Germany.
Phys Med Biol
December 2024
Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, TX 75390, United States of America.
This study analyzed the spectral response of EBT3, EBT4, and EBT-XD radiochromic films using absorption spectroscopy. The primary focus was on characterizing the evolution of spectral signatures across a range of absorbed doses, thereby elucidating the unique dose-dependent response profiles of each film type. Ten samples of each film type were subjected to open field irradiation within their designated dose ranges (1-20 Gy for EBT3 and EBT4, 1-50 Gy for EBT-XD).
View Article and Find Full Text PDFJ Contemp Brachytherapy
August 2024
Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.
Sci Rep
November 2024
Department of Experimental and Clinical Biomedical Sciences "Mario Serio", University of Florence, Florence, Italy.
The sensitivity of radiochromic films to UV-blue light is increasingly considered for light dosimetry purposes, owing to their bidimensional detection capabilities and ease of use. While film response to radiation intensity has been widely investigated by commercial scanners, spatial resolution studies remain scarce, especially for small field-of-view applications. These are of growing interest due to the antimicrobial or photo-bio-stimulating effects of UV-blue light sources in in vitro, ex vivo and in vivo models, where precise knowledge of irradiation conditions with adequate spatial resolution is crucial.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!