  • The study validated a software tool (Stroke Risk Analysis®) for detecting atrial fibrillation (AF) using ECG data from sleep studies (polysomnography).
  • The automated analysis identified AF and paroxysmal AF with varying sensitivity and specificity compared to visual analysis by a cardiologist.
  • The findings suggest that sleep studies can provide useful ECG data for AF assessment, but both the software and visual methods have low sensitivity, indicating the need for repeated ECG recordings.

Article Abstract

Purpose: The present study validated a software-based electrocardiogram (ECG) analysis tool for detection of atrial fibrillation (AF) and risk for AF using polysomnography (PSG)-derived ECG recordings.

Methods: The Stroke Risk Analysis® (SRA®) software was applied to 3-channel ECG tracings from diagnostic PSG performed in enrolled subjects including a subgroup of subjects with previously documented AF. No subjects used positive airway pressure therapy. All ECG recordings were visually analyzed by a blinded cardiologist.

Results: Of subjects enrolled in the study, 93 had previously documented AF and 178 of 186 had an ECG that could be analyzed by either method. In subjects with known history of AF, automated analysis using SRA® classified 47 out of 87 ECG as either manifest AF or showing increased risk for paroxysmal AF (PAF) by SRA® (sensitivity 0.54, specificity 0.86). On visual analysis, 36/87 ECG showed manifest AF and 51/87 showed sinus rhythm. Among the latter subgroup, an increased risk for PAF was ascribed by SRA® in 11 cases (sensitivity 0.22, specificity 0.78) and by expert visual analysis in 5 cases (sensitivity 0.1, specificity 0.90). Among 36/178 ECG with manifest AF on visual analysis, 33 were correctly identified by the SRA® software (sensitivity and specificity 0.92).

Conclusion: Sleep studies provide a valuable source of ECG recordings that can be easily subjected to software-based analysis in order to identify manifest AF and automatically assess the risk of PAF. For optimal evaluability of data, multiple channel ECG tracings are desirable. For assessment of PAF risk, the SRA® analysis probably excels visual analysis, but sensitivity of both methods is low, reflecting that repeated ECG recording remains essential.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10539451PMC
http://dx.doi.org/10.1007/s11325-023-02779-3DOI Listing

Publication Analysis

Top Keywords

visual analysis
16
ecg manifest
12
ecg
11
atrial fibrillation
8
analysis
8
sra® software
8
ecg tracings
8
ecg recordings
8
increased risk
8
risk paf
8

Similar Publications

Association between remnant cholesterol (RC) and endometriosis: a cross-sectional study based on NHANES data.

Lipids Health Dis

January 2025

Department of Obstetrics and Gynecology, Center for Reproductive Medicine, Guangdong Provincial Key Laboratory of Major Obstetric Diseases, Guangdong Provincial Clinical Research Center for Obstetrics and Gynecology, Guangdong-Hong Kong-Macao Greater Bay Area Higher Education Joint Laboratory of Maternal-Fetal Medicine, The Third Affiliated Hospital, Guangzhou Medical University, Guangzhou, China.

Background: Prior research indicates a potential link between dyslipidemia and endometriosis (EMs). However, the relationship between remnant cholesterol (RC) and EMs has not been thoroughly investigated. Consequently, looking into and clarifying the connection between RC and EMs was the primary goal of this study.

View Article and Find Full Text PDF

Background: Wrist synovitis often leads to persistent pain, swelling, and limited mobility, causing significant functional impairment and psychological distress, including anxiety. Although arthroscopic synovectomy can alleviate physical symptoms, whether it also improves anxiety and the causal relationship between synovitis and anxiety remains unclear.

Methods: A prospective study was conducted on 44 patients diagnosed with wrist synovitis who underwent arthroscopic synovectomy.

View Article and Find Full Text PDF

To assess the choroidal vessels in healthy eyes using a novel three-dimensional (3D) deep learning approach. In this cross-sectional retrospective study, swept-source OCT 6 × 6 mm scans on Plex Elite 9000 device were obtained. Automated segmentation of the choroidal layer was achieved using a deep-learning ResUNet model along with a volumetric smoothing approach.

View Article and Find Full Text PDF

Through the mobilization of movable objects due to the extreme hydraulic conditions during a flood event, blockages, damage to infrastructure, and endangerment of human lives can occur. To identify potential hazards from aerial imagery and take appropriate precautions, a change detection tool (CDT) was developed and tested using a study area along the Aisch River in Germany. The focus of the CDT development was on near real-time analysis of point cloud data generated by structure from motion from aerial images of temporally separated surveys, enabling rapid and targeted implementation of measures.

View Article and Find Full Text PDF

In response to the rotary ploughing equipment in the stubble land to implement protective operations, the stubble is large in number and strong in toughness, not easy to crush, resulting in rotary ploughing equipment to produce entanglement and increased resistance to rotary ploughing and other issues. In this study, researchers designed a bionic rotary tillage blade (B-RTB) based on the bionic structural equations of the Marmota claw. A straw-soil complex shear performance test was conducted to investigate the effect of straw on soil shear strength.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!