In a series of 14 dogs, fractures of both tibiae were caused by a "bone-breaker" designed in the authors' department and observed to produce a consistent and realistic canine fracture. One tibia was plated with a carbon fiber reinforced plastic (CFRP) plate and the other with a dynamic compression (DC) plate. Roentgenographic examination demonstrated healing of the CFRP-plated tibiae with abundant callus, and almost total remodeling of the fracture callus between ten and 20 weeks. Biomechanical testing by three-point bending revealed little difference between the strength of union of the fractures at 12-16 weeks. At 20 weeks, although the numbers were too small for statistical confirmation, the CFRP-plated tibiae were consistently stronger than the DC-plated tibiae.