Objective: To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.

Design: Observational, quality improvement study.

Setting: University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.

Methods: Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.

Results: The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).

Conclusions: We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10507511PMC
http://dx.doi.org/10.1017/ice.2022.319DOI Listing

Publication Analysis

Top Keywords

risk contamination
16
endoscopic retrograde
12
retrograde cholangiopancreatography
12
endoscopic ultrasound
12
contamination
12
contamination risk
12
linear echoendoscopes
12
risk
9
cholangiopancreatography endoscopic
8
ultrasound endoscope
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!