Objectives: Liquid chromatography-tandem mass spectrometry (LC-MS/MS) has become a common technique in clinical laboratories in recent years. Because most methods are laboratory-developed tests (LDTs), their reproducibility and quality control (QC) have been controversial. In this study, Westgard Sigma Rules were used to evaluate the analytical performance and establish an individualised internal QC (IQC) strategy for these LDTs.

Methods: Taking the LC-MS/MS LDT method for homocysteine (Hcy) as an example, the 'desirable specifications' from the Biological Variation Database were used as quality goals. Based on the external quality assessment (EQA) samples, bias was calculated and the coefficient of variation (CV) was also calculated by IQC measurements for six consecutive months. The analytical performance was evaluated by calculated sigma metrics and an IQC strategy was designed using the Westgard Sigma Rules with run size.

Results: Over 116 days within 6 months, a total of 850 data points were collected for each of IQC 1 and IQC 2. The monthly coefficient of variation CV% was 2.57-4.01%, which was non-significant (p-value: 0.75). The absolute bias% for IQC1 and IQC2 was 1.23 and 1.87%, respectively. The allowable total error (TEa) was selected as 15.5%, Sigma metrics were 4.02 and 4.30, and the analytical performance was 'Good'. The 1/2/R/4 multi rules (n=4, r=1) with a run size of 200 samples were suggested for the Hcy IQC scheme. The quality goal index (QGI) values were over 1.2, indicating that trueness needed to be improved.

Conclusions: The analytical performance of the Hcy LC-MS/MS LDT conformed to the Six Sigma rating level, achieving 'good' (four Sigma). Clinical practice indicated that calibration bias was the primary factor affecting trueness.

Download full-text PDF

Source
http://dx.doi.org/10.1515/cclm-2022-0805DOI Listing

Publication Analysis

Top Keywords

analytical performance
20
quality control
8
westgard sigma
8
sigma rules
8
iqc strategy
8
lc-ms/ms ldt
8
coefficient variation
8
sigma metrics
8
sigma
6
iqc
6

Similar Publications

A simple, rapid, and reproducible high-performance liquid chromatography (HPLC) method has been developed and validated for the determination of β-sitosterol in the pharmaceutical dosage form of moist exposed burn ointment (MEBO). This method involved an effective sample procedure for extraction of β-sitosterol from MEBO using an alkali saponification agent composed of 0.8 N ethanolic NaOH and diethyl ether.

View Article and Find Full Text PDF

%diag_test: a generic SAS macro for evaluating diagnostic accuracy measures for multiple diagnostic tests.

BMC Med Inform Decis Mak

January 2025

Institute of Mathematical Sciences Centre for Health Analytics and Modelling (CHaM), Strathmore University, Nairobi, Kenya.

Background: Measures of diagnostic test accuracy provide evidence of how well a test correctly identifies or rules-out disease. Commonly used diagnostic accuracy measures (DAMs) include sensitivity and specificity, predictive values, likelihood ratios, area under the receiver operator characteristic curve (AUROC), area under precision-recall curves (AUPRC), diagnostic effectiveness (accuracy), disease prevalence, and diagnostic odds ratio (DOR) etc. Most available analysis tools perform accuracy testing for a single diagnostic test using summarized data.

View Article and Find Full Text PDF

Low-temperature catalytic oxidation of ethanol over doped nickel phosphates.

Environ Sci Pollut Res Int

January 2025

Laboratory of Coordination and Analytical Chemistry (LCCA), Department of Chemistry, Faculty of Sciences, Chouaïb Doukkali University, Ben Maachou Road, B.P: 20, 24000, El Jadida, Morocco.

This work is focused on the synthesis and performance of Ni(PO)-based catalysts doped with Cu, Co, Mn, Ce, Zr, and Mg for the complete oxidation of ethanol, aiming at reducing emissions from ethanol-blended gasoline. Nickel phosphate was prepared via the co-precipitation method, followed by impregnation with the specified dopants. The catalysts were thoroughly characterized by XRD, N-physisorption, XRF, FTIR and Raman spectroscopy, FESEM, NH-TPD, CO-TPD, and H-TPR to explain their performance.

View Article and Find Full Text PDF

Design and validation of a simulation-based training module for ileo-transverse intracorporeal anastomosis.

Surg Endosc

January 2025

Colorectal Surgery Unit, Department of Digestive Surgery, Pontificia Universidad Católica de Chile, Uc-Christus Health Network, Santiago, Chile.

Background: The benefits of the totally laparoscopic right hemicolectomy have been established, but its adoption has been limited by the challenges of intracorporeal suturing. While simulation is effective for training advanced surgical skills, no dedicated simulation-based course exists for intracorporeal ileo-transverse anastomosis (ICA). This study aimed to develop and validate a simulation module for training in ICA.

View Article and Find Full Text PDF

Toward efficient slide-level grading of liver biopsy via explainable deep learning framework.

Med Biol Eng Comput

January 2025

Pathology Department, Beijing Youan Hospital, Capital Medical University, Beijing, 100000, China.

In the context of chronic liver diseases, where variability in progression necessitates early and precise diagnosis, this study addresses the limitations of traditional histological analysis and the shortcomings of existing deep learning approaches. A novel patch-level classification model employing multi-scale feature extraction and fusion was developed to enhance the grading accuracy and interpretability of liver biopsies, analyzing 1322 cases across various staining methods. The study also introduces a slide-level aggregation framework, comparing different diagnostic models, to efficiently integrate local histological information.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!