Background: With the clinical release of a photon counting detector-based computed tomography (CT) system, the potential benefits of this new technology need to be evaluated clinically. Literature concerning this new generation of detector is sparse, especially in the field of pediatric radiology. Therefore, this study outlines our initial experience with ultra-low dose chest CT imaging on the new photon counting CT system.
Materials And Methods: A pediatric phantom (1-year old, CIRS ATOM phantom, model 704 [CIRS-computerized imaging reference system, Norfolk, VA]) was scanned at different dose levels and different image quality levels to define a protocol for clinical examinations. Next, 20 consecutive pediatric non-contrast ultra-low dose chest CT examinations were evaluated for radiation dose and diagnostic image quality using a 4-point Likert-scale-1 = excellent, 4 = bad image quality-by two radiologists in a consensus reading. This retrospective analysis was approved by the local research ethics committee.
Results: Chest CT examinations performed at ultra-low radiation dose (effective dose 0.19 ± 0.07 mSv; size-specific dose estimate 0.45 ± 0.14 mGy) in pediatric patients ages (2.6 ± 1.8 years) show good to excellent image quality for lung structures (1.4 ± 0.4) and moderate image quality for soft tissue structures (2.8 ± 0.2).
Conclusion: Pediatric ultra-low dose chest CT examinations are feasible with the new generation photon counting detector-based CT system. The benefits of this technology must be evaluated for pediatric patients from the outset.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10212801 | PMC |
http://dx.doi.org/10.1007/s00247-022-05584-4 | DOI Listing |
Appl Radiat Isot
January 2025
Internal Dosimetry Section, Radiation Safety Systems Division, Bhabha Atomic Research Centre, Trombay, 400085, India.
Shadow Shield Whole Body Counter (SSWBC) is used to estimate internal dose of radiation workers due to the intake of fission and activation products. The SSWBC geometry was numerically modelled in FLUKA code. The computational model was validated by comparing the experimental and simulated counting efficiencies (CEs), also known as response, using Bhabha Atomic Research Centre (BARC) reference BOttle Mannequin Absorption (BOMAB) phantom.
View Article and Find Full Text PDFRadiology
January 2025
From the Department of Radiology, Duke University Hospital, 2301 Erwin Rd, Box 3808, Durham, NC 27701 (B.W.T., K.R.K., B.C.A., S.P.T., D.E.K., B.H., M.R.B., D.M., E.S., E.A.); Department of Biostatistics and Bioinformatics (N.F., S.M., A.E.) and Department of Medical Physics (W.P.S., E.S., E.A.), Duke University, Durham, NC.
Background Detection of hepatic metastases at CT is a daily task in radiology departments that influences medical and surgical treatment strategies for oncology patients. Purpose To compare simulated photon-counting CT (PCCT) with energy-integrating detector (EID) CT for the detection of small liver lesions. Materials and Methods In this reader study (July to December 2023), a virtual imaging framework was used with 50 anthropomorphic phantoms and 183 generated liver lesions (one to six lesions per phantom, 0.
View Article and Find Full Text PDFRadiology
January 2025
From the Department of Radiology, Gustave Roussy Cancer Campus, 114 Rue Edouard Vaillant, 94805 Villejuif, France.
Diagnostics (Basel)
January 2025
Clinic for Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Augsburg, Stenglinstr. 2, 86156 Augsburg, Germany.
: The number of incidental renal lesions identified in CT scans of the abdomen is increasing. Objective: The aim of this study was to determine whether hyperdense renal lesions without solid components in a portal venous CT scan can be clearly classified as vascular or non-vascular by material decomposition into iodine and water. This retrospective single-center study included 26 patients (mean age 72 years ± 9; 16 male) with 42 hyperdense renal lesions (>20 HU) in a contrast-enhanced Photon-Counting Detector CT scan (PCD-CT) between May and December 2022.
View Article and Find Full Text PDFJ Comput Assist Tomogr
January 2025
GE HealthCare, Waukesha, WI.
Objective: Patient positioning during clinical practice can be challenging, and mispositioning leads to a change in CT number. CT number fluctuation was assessed in single-energy (SE) EID, dual-energy (DE) EID, and deep silicon photon-counting detector (PCD) CT over water-equivalent diameter (WED) with different mispositions.
Methods: A phantom containing five clinically relevant inserts (Mercury Phantom, Gammex) was scanned on a clinical EID CT and a deep silicon PCD CT prototype at vertical positions of 0, 4, 8, and 12 cm.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!