Objective: To calculate the positive likelihood ratio to determine whether telemedicine is able to optimize referral to the emergency department.

Methods: Unicenter study with 182 consecutive patients admitted to Hospital Israelita Albert Einstein due to respiratory symptoms. All patients were submitted to oxygen saturation measurement using the standard method Welch Allyn finger device vital sign monitor and a 2-minute evaluation (Binah.ai mobile application). The reproducibility of oxygen saturation measurements made with both methods was investigated using interclass correlation coefficients and analysis of dispersion. Bland-Altman plots were constructed and kappa concordance coefficients used to examine data normality. Accuracy was also estimated.

Results: Oxygen saturation measurement differences between methods were ≤2% in more than 85% of cases. The mean difference (bias) between methods was near zero (0.835; Bland-Altman analysis). Oxygen saturation measurements made using the Binah.ai mobile application had an average ability to detect patients with altered oxygen saturation levels compared to the conventional method (ROC analysis). The positive likelihood ratio of the mobile application was 6.23.

Conclusion: Mobile applications for oxygen saturation measurement are accessible user-friendly tools with moderate impact on clinical telemedicine evaluation of patients with respiratory symptoms, and may optimize referral to the emergency department.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9785575PMC
http://dx.doi.org/10.31744/einstein_journal/2023AO0025DOI Listing

Publication Analysis

Top Keywords

oxygen saturation
24
respiratory symptoms
12
saturation measurement
12
mobile application
12
patients respiratory
8
positive likelihood
8
likelihood ratio
8
optimize referral
8
referral emergency
8
binahai mobile
8

Similar Publications

Background: Anthrax is a life-threatening zoonotic disease caused by Gram-positive, spore-forming bacterium . It manifests as a cutaneous, gastrointestinal, and respiratory disease. The cutaneous form ranges from a self-limiting lesion to severe edematous lesions with toxemic shock.

View Article and Find Full Text PDF

A 69-year-old man with chest pain was diagnosed with acute type B aortic dissection with the entry tear located at distal arch and a distal aortic arch aneurysm. Therefore, we performed debranching thoracic endovascular aortic repair 2 weeks after type B aortic dissection onset. First, the graft was anastomosed to bilateral axillary arteries.

View Article and Find Full Text PDF

The risk of cardiovascular disease differs among various ethnic groups, highlighting disparities in cardiovascular health among different populations. While multiple studies from other countries have looked at changes in physiological parameters during autonomic function tests like isometric handgrip and cold pressor tests, no correlational research has been done in Saudi Arabia. This lacuna underscores the importance of examining the relationship between cardiorespiratory parameters in young Saudi Arabian individuals during these tests.

View Article and Find Full Text PDF

Background: This study investigates the correlation between the difference in hematocrit (HCT) and serum albumin (ALB) levels (HCT-ALB), muscle tissue oxygen saturation (SmtO2), capillary refill time (CRT), and blood lactate (Lac) with the severity of renal function damage and prognosis in patients with septic shock.

Methods: Conducted from February 2022 to February 2024, this study included 116 septic shock patients treated at the First Affiliated Hospital of Anhui Medical University. Patients were divided into groups based on whether they developed acute kidney injury: 40 patients were included in the acute kidney injury group, and the remaining 76 were placed in the non-kidney injury group.

View Article and Find Full Text PDF

Photoacoustic microscopy has demonstrated outstanding performance in high-resolution functional imaging. However, in the process of photoacoustic imaging, the photoacoustic signals will be polluted by inevitable background noise. Besides, the image quality is compromised due to the biosafety limitation of the laser.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!