Background/purpose: pCONUS 2 and pCONUS 2-HPC are neck-bridging devices that provide coiling support in the endovascular treatment of wide-necked intracranial aneurysms. To date, limited multicentre data has been published. This study provides the first pooled data from multiple UK centres regarding outcomes for these devices covering the periprocedural period to 6-month follow-up.
Materials/methods: This retrospective, single-arm study assessed 65 patients treated over 3 years from the time of procedure to 6 months post-procedure across four UK centres. Data collected included patient demographics, aneurysm characteristics and antiplatelet regimens. Outcome measures were angiographic results and procedure-related complications from the immediate periprocedural period to 6-month follow-up.
Results: Fifty-four unruptured (83.1%) and 11 ruptured (16.9%) aneurysms were treated. Fifty-five aneurysms were located in the anterior circulation (87.7%). There were four device-related intraprocedural complications: three cases of asymptomatic, temporary thrombus formation and one mortality associated with branch vessel occlusion and aneurysm re-bleeding in a ruptured case. There were no post-procedural device-related complications. Satisfactory occlusion was achieved in 58/65 procedures (89.2%) at time of treatment and in 44/60 (73.3%) at 6 months. Satisfactory occlusion correlated with aneurysm size and coiling packing density. Retreatment was required for five unruptured cases (7.7%) and was straightforward with the device in situ.
Conclusion: pCONUS 2 and pCONUS 2-HPC have good short-term safety profiles. The use of pCONUS 2-HPC in the acute treatment of ruptured aneurysms with postprocedural SAPT is feasible. The devices have an intraprocedural complication rate of 4/65 (6.2%) across multiple UK centres, including a single death (1.5%).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1177/15910199221150467 | DOI Listing |
Interv Neuroradiol
January 2023
Radiology Department, University Hospital of Wales, Cardiff and Vale University Health Board, Cardiff, Wales, UK.
Background/purpose: pCONUS 2 and pCONUS 2-HPC are neck-bridging devices that provide coiling support in the endovascular treatment of wide-necked intracranial aneurysms. To date, limited multicentre data has been published. This study provides the first pooled data from multiple UK centres regarding outcomes for these devices covering the periprocedural period to 6-month follow-up.
View Article and Find Full Text PDFWorld Neurosurg
July 2021
Radiology Departments, University Hospital Saint Ivan Rilski, Sofia, Bulgaria. Electronic address:
A variety of recently introduced devices have allowed more complex aneurysms to be rendered for endovascular embolization. Tiny and sophisticated implants are nowadays designed to provide further either temporary or permanent remodeling of the wide-necked aneurysms and improve coil occlusion rates. Such improvements in the technical armamentarium include the Woven EndoBridge (Sequent Medical, California, USA), the Medina Embolic Device (Medtronic, Irvine, California, USA), The Contour Neurovascular System (Cerus Endovascular, Fremont, California, USA), and pCONUS 2 HPC (Phenox, GmbH, Bohum, Germany).
View Article and Find Full Text PDFInterv Neuroradiol
December 2021
University Hospital of Wales, Cardiff, UK.
Background/purpose: pCONUS 2 and pCONUS 2-HPC are novel neck-bridging devices designed to support endovascular coil occlusion of wide-necked intracranial bifurcation aneurysms. This study summarises periprocedural outcomes, 6-month, and early 2-year follow-up results following its introduction in an interventional neuroradiology centre.
Materials/methods: This prospective, single-arm study assessed 20 aneurysms treated over a 40 month time period from time of procedure to 2 years post-procedure.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!