Background: Reveal® 3-D for Gluten is an immunochromatographic assay for the qualitative detection of gluten in environmental samples. The test uses monoclonal antibodies reactive to prolamins in wheat.
Objective: The objective of the study was to validate the Reveal 3-D test for detection of gluten in clean-in-place rinse and swabs from a stainless steel surface.
Methods: Elements of the study included food selectivity and interference testing, matrix testing, an assay robustness study, and reagent stability/lot-to-lot consistency testing. Wheat flour was used as the spiking material for all matrixes.
Results: In selectivity and interference testing, nine target matrixes all tested positive and 36 of 39 non-target matrixes tested negative. Almond flour, sesame flour, and cornstarch produced positive results as 100% commodities; reactivity can be eliminated with dilution or by testing without use of food extraction buffer, which is not a standard part of the environmental testing method. With a gluten spike at 9.3 mg/kg, chestnut flour, guar gum, and xanthan gum as 100% commodities inhibited the ability of the assay to detect gluten when tested without dilution. In quaternary ammonium clean-in-place rinse and swabs from stainless steel, 100% positive results were obtained at levels of 2.8 mg/kg and 4.7 µg/100 cm2, respectively. Results of independent laboratory testing of swabs from stainless steel supported those of internal trials. Robustness testing showed that introducing variations to three operating parameters simultaneously had no adverse effect on assay performance. In the reagent stability study, data supported kit expiration dating of 11 months.
Conclusion: Results of the current study show that the Reveal test is an accurate and reliable method for qualitative detection of gluten in select clean-in-place rinse and environmental samples.
Highlights: The Reveal test was able to detect gluten at levels of 2.8 ppm in clean-in-place rinse and 4.7 µg/100 cm2 in swabs from stainless steel.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/jaoacint/qsac157 | DOI Listing |
Nanomaterials (Basel)
December 2024
School of Intelligent Manufacturing, Luoyang Institute of Science and Technology, Luoyang 471023, China.
(AlCrMoNiTi)N high-entropy alloy nitride (HEAN) films were synthesized at various bias voltages using the co-filter cathodic vacuum arc (co-FCVA) deposition technique. This study systematically investigates the effect of bias voltage on the microstructure and performance of HEAN films. The results indicate that an increase in bias voltage enhances the energy of ions while concomitantly reducing the deposition rate.
View Article and Find Full Text PDFInt J Surg
December 2024
Department of Urology, Zhongnan Hospital of Wuhan University, Wuhan, China.
Purpose: To evaluate the effectiveness and safety of an upgraded integrated vacuum suction catheter in semi-rigid ureteroscopic laser lithotripsy (VC-URSL) compared to traditional methods for treating impacted upper ureteral stones.
Patients And Methods: This prospective, randomized controlled trial was conducted from September 2022 to March 2024 at a single center, enrolling 95 patients aged 18 to 70 years with a single radiopaque impacted upper ureteral stone. Participants were randomized into two groups: the VC-URSL group used an integrated vacuum suction catheter featuring a stainless steel stabilizing tube and a narrowed distal end to prevent obstruction, while the T-URSL group underwent standard ureteroscopic lithotripsy without vacuum assistance.
J Funct Biomater
November 2024
King Abdullah International Medical Research Center, Riyadh 11481, Saudi Arabia.
The aim of the study is to assess the impact of mechanical surface treatments on the shear bond strength (SBS) of orthodontic brackets bonded to three-dimensional (3D) printed and milled CAD/CAM provisional materials. Sixty cylindrical samples were fabricated for each provisional material. Samples were treated with one of the following surface treatments: aluminum oxide airborne particle abrasion, diamond bur rotary instrument roughening, and phosphoric acid etching (control).
View Article and Find Full Text PDFDent J (Basel)
December 2024
Research Center for Digital Technologies in Dentistry and CAD/CAM, Department of Dentistry, Faculty of Medicine and Dentistry, Danube Private University, Steiner Landstraße 124, 3500 Krems an der Donau, Austria.
: Traditional methods for palatal expansion using fixed appliances often face limitations in comfort and aesthetics. In comparison, aligner therapy has limitations, particularly regarding maxillary expansion. The aim of this study is to examine the biomechanical properties regarding the wire diameter and bending of different stainless steel wires to evaluate their potential for incorporation into maxillary aligner therapy.
View Article and Find Full Text PDFCureus
November 2024
Faculty of health and medical technology, Department of Medical Laboratory Technology, Al-Ayen University, Nasiriyah, IRQ.
Background Successful endodontic treatment depends on several elements. The most important of which is the elimination of microorganisms within the root canal system. To achieve this, sterility must be maintained during all steps of the treatment.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!