Purpose: Ulnar shortening osteotomy can be used to treat ulnar impaction syndrome and other causes of ulnar wrist pain. Distal metaphyseal ulnar shortening osteotomy (DMUSO) is one technique that has been proposed to reduce the complications seen with a diaphyseal USO or a wafer resection. However, to our knowledge, the optimal fixation construct for DMUSO has not been studied. We sought to characterize the biomechanical stiffness and rotational stability of different DMUSO constructs.

Methods: A DMUSO was performed on 40 human cadaveric ulnas using 4 different fixation constructs (10 specimens per group): one 3.0 mm antegrade screw; two 2.2 mm antegrade screws; one 3.0 mm retrograde screw; and two 2.2 mm retrograde screws. Biaxial testing using axial load and cyclical axial torque was performed until failure, defined as 10° of rotation or 2 mm displacement. Specimens were assessed for stiffness at failure. Bone density was assessed using the second metacarpal cortical percentage.

Results: Bone density was similar between all 4 testing groups. Of the 4 groups, the 2 antegrade screw group exhibited the highest rotational stiffness of 232 ± 102 Nm/deg. In paired analysis, this was significantly greater than 1 retrograde screw constructs. In multivariable analysis, 2-screw constructs were significantly stiffer than 1 screw and antegrade constructs were significantly stiffer than retrograde. Maximum failure torque did not differ with orientation, but 2 screws failed at significantly higher torques.

Conclusion: Using 2 screws for DMUSO fixation constructs may provide higher stiffness and maximum failure torque, and antegrade screw constructs may provide more stiffness than retrograde constructs.

Clinical Relevance: Antegrade screw fixation using 2 screws may provide the strongest construct for DMUSO. Antegrade fixation may be preferred because it avoids violating the distal radioulnar joint capsule and articular surface of the ulna.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jhsa.2022.11.007DOI Listing

Publication Analysis

Top Keywords

antegrade screw
16
ulnar shortening
12
shortening osteotomy
12
distal metaphyseal
8
metaphyseal ulnar
8
construct dmuso
8
fixation constructs
8
screw antegrade
8
retrograde screw
8
bone density
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!