  • The study evaluated the relationship between diastolic pressure ratio (dPR) and fractional flow reserve (FFR) in patients with coronary artery stenosis in Japan.
  • It included 100 patients from four hospitals and found a strong correlation (R = 0.841) between dPR and FFR, with diagnostic agreement in 89% of cases.
  • The study highlighted that diagnostic differences were influenced by factors like the location of the lesions and the medication used, and noted a slightly higher concordance rate when using adenosine compared to nicorandil.

Article Abstract

Purpose: This study aimed to evaluate the correlation and diagnostic agreement between diastolic pressure ratio (dPR) and fractional flow reserve (FFR) in a Japanese real-world setting.

Design: Prospective multicenter observational study.

Methods: This study included 100 patients with intermediate coronary artery stenosis at 4 Japanese hospitals. For these lesions, FFR and dPR were measured using a guidewire with a sensor and a monitor to measure intravascular pressure. The correlation and diagnostic agreement between FFR and dPR were assessed. When both FFR and dPR were negative or positive, the results were considered to be concordant. When one was positive and the other was negative, the result was regarded as discordant (positive discordance, FFR > 0.80 and dPR ≤ 0.89; negative discordance, FFR ≤ 0.80 and dPR > 0.89).

Results: Overall, the FFR and dPR were well-correlated (R = 0.841). FFR and dPR were concordant in 89% of cases (concordant normal, 43%; concordant abnormal, 46%) and discordant in 11% (positive discordance, 7%; negative discordance, 4%). No significant difference was observed in the rate of concordant results between patients with and without diabetes mellitus. The diagnostic concordance rate was significantly different among the 3 coronary arteries (right coronary artery, 93.3%; left anterior descending artery, 93.2%; and left circumflex artery, 58.3%; P = .001). Additionally, the rate of concordant results tended to be higher when using intravenous administration of adenosine than when using intracoronary bolus injection of nicorandil (adenosine, 95.1%; nicorandil, 84.7%; P = .103).

Conclusion: We found that dPR was highly correlated with FFR, and diagnostic discordance was observed in 11% of the lesions. Several factors, including lesion location and medication for hyperemia, may cause the diagnostic discordance between dPR and FFR.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9803508PMC
http://dx.doi.org/10.1097/MD.0000000000032578DOI Listing

Publication Analysis

Top Keywords

ffr dpr
20
fractional flow
8
flow reserve
8
diastolic pressure
8
pressure ratio
8
guidewire sensor
8
intravascular pressure
8
correlation diagnostic
8
diagnostic agreement
8
dpr
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!