Comprehensive comparisons of ocular biometry: A network-based big data analysis.

Eye Vis (Lond)

Eye Institute and Department of Ophthalmology, Institute for Medical and Engineering Innovation, Eye & ENT Hospital, Fudan University; NHC Key Laboratory of Myopia (Fudan University), Key Laboratory of Myopia, Chinese Academy of Medical Sciences, No. 19 Baoqing Road, Xuhui District, Shanghai, 200031, China.

Published: December 2022

Purpose: To systematically compare and rank ocular measurements with optical and ultrasound biometers based on big data.

Methods: PubMed, Embase, the Cochrane Library and the US trial registry ( www.

Clinicaltrial: gov ) were used to systematically search trials published up to October 22nd, 2020. We included comparative studies reporting the following parameters measured by at least two devices: axial length (AL), flattest meridian keratometry (Kf), steepest meridian keratometry (Ks), mean keratometry (Km), astigmatism (AST), astigmatism vectors J and J, anterior chamber depth (ACD), aqueous depth (AQD), central corneal thickness (CCT), corneal diameter (CD) and lens thickness (LT). A network-based big data analysis was conducted using STATA version 13.1.

Results: Across 129 studies involving 17,181 eyes, 12 optical biometers and two ultrasound biometers (with both contact and immersion techniques) were identified. A network meta-analysis for AL and ACD measurements found that statistically significant differences existed when contact ultrasound biometry was compared with the optical biometers. There were no statistically significant differences among the four swept-source optical coherence tomography (SS-OCT) based devices (IOLMaster 700, OA-2000, Argos and ANTERION). As for Ks, Km and CD, statistically significant differences were found when the Pentacam AXL was compared with the IOLMaster and IOLMaster 500. There were statistically significant differences for CCT when the OA-2000 was compared to Pentacam AXL, IOLMaster 700, Lenstar, AL-Scan and Galilei G6.

Conclusion: For AL and ACD, contact ultrasound biometry obtains the lower values compared with optical biometers. The Pentacam AXL achieves the lowest values for keratometry and CD. The smallest value for CCT measurement is found with the OA-2000.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9808957PMC
http://dx.doi.org/10.1186/s40662-022-00320-3DOI Listing

Publication Analysis

Top Keywords

statistically differences
16
optical biometers
12
pentacam axl
12
meridian keratometry
8
contact ultrasound
8
ultrasound biometry
8
compared optical
8
iolmaster 700
8
optical
5
comprehensive comparisons
4