Background/aim: The aim was to evaluate the effect of a combined precision matrix and high sampling rate on the delineation of anatomical structures and objective image quality in single source CT in a qualitative approach.

Materials And Methods: An anthropomorphic thoracic phantom was used to evaluate the objective image quality parameters, including image noise, noise power spectrum, image stepness and Q for different CT scanners including high/standard matrix and framing frequency setups. Scan parameters were standardized over all scanners. Additional subjective quality assessment was also performed.

Results: A linear mixed effects model was used to determine the effect of sampling rate and image matrix on objective image quality parameters. Noise power spectrum and image noise were significantly influenced by both framing frequency and image matrix. There were significant differences between high and standard frequency/matrix acquisitions.

Conclusion: Higher framing frequency and image matrix allows for improved image noise texture and objective image quality in CT.