Purpose: To investigate the feasibility of ultra-low-dose CT (ULDCT) reconstructed with the artificial intelligence iterative reconstruction (AIIR) algorithm in total-body PET/CT imaging.

Methods: The study included both the phantom and clinical parts. An anthropomorphic phantom underwent CT imaging with ULDCT (10mAs) and standard-dose CT (SDCT) (120mAs), respectively. ULDCT was reconstructed with AIIR and hybrid iterative reconstruction (HIR) (expressed as ULDCT-AIIR and ULDCT-HIR), respectively, and SDCT was reconstructed with HIR (SDCT-HIR) as control. In the clinical part, 52 patients with malignant tumors underwent the total-body PET/CT scan. ULDCT with AIIR (ULDCT-AIIR) and HIR (ULDCT-HIR), respectively, was reconstructed for PET attenuation correction, followed by the SDCT reconstructed with HIR (SDCT-HIR) for anatomical location. PET/CT images' quality was qualitatively assessed by two readers. The CT, as well as the CT standard deviation (CT), SUV, SUV, and the SUV standard deviation (SUV), was recorded. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated and compared.

Results: The image quality of ULDCT-HIR was inferior to the SDCT-HIR, but no significant difference was found between the ULDCT-AIIR and SDCT-HIR. The subjective score of ULDCT-AIIR in the neck, chest and lower limb was equivalent to that of SDCT-HIR. Besides the brain and lower limb, the change rates of CT in thyroid, neck muscle, lung, mediastinum, back muscle, liver, lumbar muscle, first lumbar spine and sigmoid colon were -2.15, -1.52, 0.66, 2.97, 0.23, 8.91, 0.06, -4.29 and 8.78%, respectively, while all CT of ULDCT-AIIR was lower than that of SDCT-HIR. Except for the brain, the CNR of ULDCT-AIIR was the same as the SDCT-HIR, but the SNR was higher. The change rates of SUV, SUV and SUV were within [Formula: see text] 3% in all ROIs. For the lesions, the SUV, SUV and TBR showed no significant difference between PET-AIIR and PET-HIR.

Conclusion: The SDCT-HIR could not be replaced by the ULDCT-AIIR at date, but the AIIR algorithm decreased the image noise and increased the SNR, which can be implemented under special circumstances in PET/CT examination.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9807709PMC
http://dx.doi.org/10.1186/s40658-022-00521-8DOI Listing

Publication Analysis

Top Keywords

suv suv
20
iterative reconstruction
12
total-body pet/ct
12
suv
9
reconstructed artificial
8
artificial intelligence
8
intelligence iterative
8
pet/ct examination
8
uldct reconstructed
8
aiir algorithm
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!