Background: Phase-based electrical property tomography (EPT) is a technique that allows conductivity reconstruction with only phase of the B field under the assumption that the magnitude of the B fields are homogeneous. The more this assumption is violated, the less accurate the reconstructed conductivity. Thus, a method that ensures homogeneity of field is important for breast image using multi-receiver coil.

Purpose: To develop a method for multi-receiver combination for phase-based EPT usable for breast EPT imaging in the clinic.

Methods: Theory of the proposed method is presented. To validate the proposed method, the phantom and in-vivo experiments were conducted. Conductivity images were reconstructed using the transceive phase of the combined image and results were compared with another combination method.

Results: The proposed method's conductivity results were more stable than those of the previous method when was not homogeneous and when the homogeneous contrast region was small. The phantom and in-vivo results indicate that the proposed method produces improved conductivity images than the previous method. The proposed combination method also increased the conductivity contrast between benign and cancerous tissues.

Conclusion: The proposed method produced more stable multi-receiver combination for phase-based EPT of the breast in a clinical environment.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16195DOI Listing

Publication Analysis

Top Keywords

proposed method
16
multi-receiver combination
12
method
10
combination method
8
phase-based electrical
8
electrical property
8
property tomography
8
combination phase-based
8
phase-based ept
8
phantom in-vivo
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!