Hypochlorous acid (HClO) plays an important role in many physiological and pathological activities. In this work, a novel BODIPY-based Near-infrared (NIR) ratiometric fluorescent probe BODIPY-Hyp was designed for the rapid detection of HClO. The probe BODIPY-Hyp was highly selective and sensitive for HClO with a low detection limit of 16.74 nM and short response time of less than 60 s. The probe BODIPY-Hyp in response to HClO exhibited a significant blue-shifted fluorescence emission from 700 nm to 530 nm, and its fluorescence intensity ratio (I/I) increased about 1200 times before and after adding HClO. Moreover, the reaction mechanism of BODIPY-Hyp with HClO was verified by HRMS analysis, H NMR titration and DFT calculations. Furthermore, BODIPY-Hyp was successfully processed into a portable test strip-based device for the detection of HClO. In addition, the probe BODIPY-Hyp could be used in real time to monitor the levels of HClO in living zebrafish larvae. In conclusion, BODIPY-Hyp has great application potential in the life and environmental sciences.