Reliability and agreement of apparent chord mu measurements between static and dynamic evaluations.

J Cataract Refract Surg

From the Instituto de Oftalmobiología Aplicada (IOBA), Universidad de Valladolid, Valladolid, Spain (Martínez-Plaza, Marcos, López-de la Rosa, López-Miguel, Maldonado); Red Temática de Investigación Colaborativa en Oftalmología (OftaRed), Instituto de Salud Carlos III, Madrid, Spain (Martínez-Plaza, López-Miguel, Maldonado).

Published: January 2023

Purpose: To assess the repeatability and agreement of Cartesian coordinates and the length of apparent chord mu and pupil diameter measurements during static (Galilei G4) and dynamic (Topolyzer Vario) evaluations.

Setting: IOBA-Eye Institute, Valladolid, Spain.

Design: Case series.

Methods: 3 consecutive measurements per scenario (Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions) were performed by the same clinician. The intrasession repeatability was assessed using the within-subject SD (Sw), the precision, the coefficient of variation, and the intraclass correlation coefficient (ICC). The agreement was analyzed using repeated-measures analysis of variance and the Bland-Altman method.

Results: Thirty-seven healthy participants were recruited. The Sw values for chord mu parameters and pupil diameter ranged from 0.01 to 0.03 and 0.08 to 0.21, respectively. The ICC was ≥0.89 for all parameters. Galilei G4 and Topolyzer Vario under low mesopic and photopic conditions provided significantly different measures of apparent chord mu length (0.23 ± 0.11 mm, 0.30 ± 0.10 mm, and 0.25 ± 0.11 mm, respectively, P ≤ .02), X-coordinate (-0.18 ± 0.12 mm, -0.27 ± 0.11 mm, and -0.21 ± 0.12 mm, respectively, P < .001), and pupil diameter (3.38 ± 0.50 mm, 6.29 ± 0.60 mm, and 3.04 ± 0.41 mm, respectively, P < .001). Y-coordinate values obtained by Galilei G4 and Topolyzer Vario under low mesopic conditions were significantly different (0.06 ± 0.13 mm vs 0.03 ± 0.11 mm, respectively, P = .02), in contrast to Galilei G4 and Topolyzer Vario under photopic conditions (0.05 ± 0.13 mm, P = .82) and both illumination conditions of Topolyzer Vario (P ≥ .23).

Conclusions: Galilei G4 and Topolyzer Vario provide consistent measurements of apparent chord mu Cartesian coordinates and length, as well as pupil diameter; however, the measurements are not interchangeable. Ophthalmic surgeons should consider these findings when planning customized intraocular lens implantation and refractive surgery procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000001036DOI Listing

Publication Analysis

Top Keywords

topolyzer vario
28
galilei topolyzer
20
apparent chord
16
pupil diameter
16
vario low
12
low mesopic
12
photopic conditions
12
measurements static
8
cartesian coordinates
8
coordinates length
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!