Assessing cortisol levels in human bodies has become essential to diagnose heart failure (HF). In this work, we propose a salivary cortisol detection strategy as part of an easily integrable lab-on-a-chip for detection of HF biomarkers. Our developed capacitive immunosensor based on hafnium oxide (HfO2)/silicon structure showed good linearity between increasing cortisol concentration and the charge-transfer resistance/capacitance. Moreover, the developed biosensor was demonstrated to be highly selective toward cortisol compared to other HF biomarkers such as tumor necrosis factor (TNF-α) and N-terminal pro-brain natriuretic peptide (NT-proBNP). The precision of our developed biosensor was evaluated, and the difference between the determined cortisol concentration in saliva and its expected one is <18%.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9786208 | PMC |
http://dx.doi.org/10.3390/mi13122235 | DOI Listing |