Screw Pullout Strength After Pedicle Screw Reposition: A Finite Element Analysis.

Spine (Phila Pa 1976)

Department of Spine Surgery, Honghui Hospital, Xi'an Jiaotong University, Xi'an, Shaanxi, China.

Published: November 2023

Research Design: Finite element analysis based on computed tomography images from the lumbar spine.

Objective: Determined the pullout strength of unsatisfactorily placed screws and repositioned screws after unsatisfactory place in lumbar spine surgery.

Background: Pedicle screws are widely used to stabilize the spinal vertebral body. Unsatisfactory screws could lead to surgical complications, and may need to be repositioned. Screw removal and reposition, however, may decrease pullout strength.

Methods: We conducted a three-dimensional finite element analysis based on high-resolution computed tomography images from a 39-year-old healthy woman. Pullout strength was determined with the screw placed in different orientations at the same entry point (as selected by the Magerl method), as well as after removal and reposition. The material properties of the vertebral body and the screw were simulated by using grayscale values and verified data, respectively. A load along the screw axis was applied to the end of the screw to simulate the pullout.

Results: The pullout strength was 1840.0 N with the Magerl method. For unsatisfactorily placed screws, the pullout strength was 1500.8 N at 20% overlap, 1609.6 N at 40% overlap, 1628.9 N at 60% overlap, and 1734.7 N at 80% overlap with the hypothetical screw path of the Magerl method. For repositioned screws, the pullout strength was 1763.6 N, with 20% overlap, 1728.3 N at 40% overlap, 1544.0 N at 60% overlap, and 1491.1 N at 80% overlap, with the original path. Comparison of repositioned screw with unsatisfactorily placed screw showed 14.04% decrease in pullout strength at 80% overlap, 5.21% decrease at 60% overlap, 7.37% increase at 40% overlap, and 17.51% increase at 20% overlap, with the screw path of the Magerl method.

Conclusions: Removal and reposition increased the pullout strength at 20% and 40% overlap, but decreased the pullout strength at 60% and 80% overlap. For clinical translation, we recommend removal and reposition of the screw when the overlap is in the range of 20% to 40% or less. In vitro specimen studies are needed to verify these preliminary findings.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10602223PMC
http://dx.doi.org/10.1097/BRS.0000000000004553DOI Listing

Publication Analysis

Top Keywords

pullout strength
36
removal reposition
16
40% overlap
16
80% overlap
16
overlap
15
screw
12
finite element
12
element analysis
12
magerl method
12
20% overlap
12

Similar Publications

Graphene and its derivatives have been widely used as reinforcing nanofillers for high-performance polymer nanocomposites. The effectiveness of the reinforcement largely depends on the properties of the nanofiller-matrix interface, which can be represented by the interfacial shear strength (IFSS). This work systematically investigates IFSS enhancements for polyethylene (PE) nanocomposites reinforced by graphene origami (GOri) through molecular dynamics pull-out simulations.

View Article and Find Full Text PDF

Introduction Spinal fusion surgery with pedicle screws is commonly performed to stabilize the spine of osteoporotic patients. However, securing a strong screw fixation in osteoporotic bone presents significant challenges due to the reduced bone density. This study aimed to compare the biomechanical performance in an osteoporotic bone model of pedicle screws inserted using two different techniques, the Jamshidi needle technique and the pedicle probe technique, as well as the influence of tapping on both these techniques.

View Article and Find Full Text PDF

Background: For complete disruption of the posterolateral corner (PLC) structures, operative treatment is most commonly advocated, as nonoperative treatment has higher rates of persistent lateral laxity and posttraumatic arthritis. Some studies have shown that acute direct repair results in revision rates upwards of 37% to 40% compared with 6% to 9% for initial reconstruction. In a recent study assessing the outcomes of acute repair of PLC avulsion injuries with 2 to 7 years of follow-up, patients with adequate tissue were shown to have a much lower failure rate than previously documented.

View Article and Find Full Text PDF

A curved compliant spinal bone anchor to enhance fixation strength.

PLoS One

December 2024

Bio-Inspired Technology Group, Faculty of Mechanical Engineering, Department of BioMechanical Engineering, Delft University of Technology, Delft, The Netherlands.

Pedicle screws have long been established as the gold standard for spinal bone fixation. However, their fixation strength can be compromised in cases of low bone density, particularly in osteoporotic bone, due to the reliance on a micro-shape lock between the screw thread and the surrounding bone. To address this challenge, we propose augmenting conventional pedicles screws with a curved compliant anchor.

View Article and Find Full Text PDF

This study presents the design and experimental evaluation of advanced corrosion protection coatings for application on prestressing strands which are the core constituents of prestressed concrete structures such as bridges. Variety of self-heal coatings embodying corrective and protective phenomena in response to the degrading effects of corrosion have been designed and tested in simulated aggressive weathering conditions. Standard 7-wire prestressing strands coated with self-heal epoxy, self-heal toughened epoxy and hybrid epoxy coating systems were subjected to salt fog spray up to a duration of 2500 h, and 3M CalCl, 3M NaOH, saturated Ca(OH) solutions and distilled water up to 45 days duration.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!