Background: A subeustachian pouch (SEP) often hinders the completion of a cavotricuspid isthmus (CTI) ablation of typical atrial flutter (AFL) and sometimes causes steam-pops during a power-controlled ablation. We hypothesized that real-time bull's-eye monitoring of the catheter surface temperature might be useful to locate the SEP where the temperature can rise rapidly, and a temperature-controlled ablation might avoid steam pops. This study aimed to demonstrate this hypothesis.
Methods: A temperature-controlled CTI ablation with a QDOT MICRO™ catheter ( = 10) and a conventional power-controlled CTI ablation ( = 10) were performed with an output power of 35 W. During the RF application, the bull's eye monitor for monitoring the catheter surface temperatures was assessed. A "red-bull sign" was defined as an entire red-colored bull's-eye monitor, indicating that the catheter-tip temperature of all 6 thermocouples rose rapidly over 47°C.
Results: In a total of 115 lesions (12 ± 3 per patient), a "red-bull sign" was observed in 39 (33.9%) lesions where the RF output was reduced to 26 ± 8 W. All 39 "red-bull sign" lesions corresponded to the location of the SEP as delineated by ICE before the ablation. The red-bull sign accurately indicated the presence of a SEP with a sensitivity of 84.7% and specificity of 100%. Bidirectional block of the CTI was completed in all patients in either catheter group without any steam-pops.
Conclusion: Real-time surface temperature monitoring and a red-bull sign might be useful to detect the SEP. A temperature-controlled CTI ablation with the QDOT MICRO catheter might be safe for avoiding steam pops.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9745479 | PMC |
http://dx.doi.org/10.1002/joa3.12793 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!