This study provides insight into the overall system performance, stability, and delivery accuracy of the first clinical self-shielded stereotactic radiosurgery (SRS) system. Quality assurance procedures specifically developed for this unit are discussed, and trends and variations over the course of 2-years for beam constancy, targeting and dose delivery are presented. Absolute dose calibration for this 2.7 MV unit is performed to deliver 1 cGy/MU at d = 7 mm at a source-to-axis-distance (SAD) of 450 mm for a 25 mm collimator. Output measurements were made with 2-setups: a device that attaches to a fixed position on the couch (daily) and a spherical phantom that attaches to the collimating wheel (monthly). Beam energy was measured using a cylindrical acrylic phantom at depths of 100 (D ) and 200 (D ) mm. Beam profiles were evaluated using Gafchromic film and compared with TPS beam data. Accuracy in beam targeting was quantified with the Winston-Lutz (WL) and end-to-end (E2E) tests. Delivery quality assurance (DQA) was performed prior to clinical treatments using Gafchromic EBT3/XD film. Net cumulative output adjustments of 15% (pre-clinical), 9% (1st year) and 3% (2nd year) were made. The mean output was 0.997 ± 0.010 cGy/MU (range: 0.960-1.046 cGy/MU) and 0.993 ± 0.029 cGy/MU (range: 0.884-1.065 cGy/MU) for measurements with the daily and monthly setups, respectively. The mean relative beam energy (D /D ) was 0.998 ± 0.004 (range: 0.991-1.006). The mean total targeting error was 0.46 ± 0.17 mm (range: 0.06-0.98 mm) for the WL and 0.52 ± 0.28 mm (range: 0.11-1.27 mm) for the E2E tests. The average gamma pass rates for DQA measurements were 99.0% and 90.5% for 2%/2 mm and 2%/1 mm gamma criteria, respectively. This SRS unit meets tolerance limits recommended by TG-135, MPPG 9a., and TG-142 with a treatment delivery accuracy similar to what is achieved by other SRS systems.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10018673PMC
http://dx.doi.org/10.1002/acm2.13857DOI Listing

Publication Analysis

Top Keywords

performance stability
8
clinical self-shielded
8
self-shielded stereotactic
8
stereotactic radiosurgery
8
delivery accuracy
8
quality assurance
8
beam energy
8
e2e tests
8
cgy/mu range
8
beam
6

Similar Publications

In this study, a walnut peptide (WP) with calcium-binding capacity was prepared using a combination of alkalase and neutrase. The conditions for the preparation of walnut peptide calcium chelate (WP-Ca) were optimised (a peptide/calcium chloride ratio of 1 : 4 for 70 min at 50 °C and pH 9.5).

View Article and Find Full Text PDF

SARS-CoV-2 Envelope (E) protein is critical in viral assembly, release, and virulence. E gene was considered highly conserved and evolving slowly. Pan-sarbecoviruses-conserved regions in the E gene have been used as targets for various RT-PCR assays to detect SARS-CoV-2.

View Article and Find Full Text PDF

Sulfur-Doped Nickel Ferrite for Green Hydrogen at High Current Density.

Chem Asian J

January 2025

Charotar University of Science and Technology, Physical Science, P.D. Patel Institute of Applied Sciences, 388421, Changa, INDIA.

The primary obstacle in electrolyzing water is that prolonged large-current operation quickly degrades performance, making it difficult to achieve efficient and continuous hydrogen evolution at high current densities. This work prepared sulfur-doped nickel ferrite nanocomposites using the simple hydrothermal method to improve electrocatalytic green hydrogen production at high-current densities. X-ray diffraction (XRD), scanning electron microscopy (SEM), and X-ray photoelectron spectroscopy (XPS) were used to analyze the crystalline structure, morphology, and chemical composition of the synthesized nanocomposites.

View Article and Find Full Text PDF

An Optoelectronic Sensing Real-Time Glucose Detection Film Using Photonic Crystal Enhanced Rare Earth Fluorescence and Additive Manufacturing.

Small

January 2025

State Key Laboratory of Electromechanical Integrated Manufacturing of High-performance Electronic Equipment, School of Mechano-Electronic Engineering, School of Life Science and Technology, Xidian University, Xi'an, Shaanxi, 710071, China.

In this research, a novel detection method employing rare-earth upconversion nanoparticle (UCNP) as the core, coated with MnO nanosheets is designed, which formed a color and fluorescence dual-responsive UCNP composite material, MnO-modified NaYF:Yb,Tm@NaYF. By enabling both colorimetric and fluorescence methods simultaneously, this composite material allows for the detection of glucose concentration under different conditions, while exhibiting strong resistance to environmental interference, chemical stability, and accuracy. To further enhance the sensitivity of the detection method, a photonic crystals (PCs)-PDMS array where polymethyl methacrylate PCs are deposited onto a substrate composed of PDMS-glass slice with hydrophobic surfaces is developed.

View Article and Find Full Text PDF

To rival commercial organic electrolytes, it is important to focus on safe, cheap aqueous electrolytes with lower salt concentration (≈5.0 m) and a wider electrochemical stable potential window (ESPW). This study reports the facile synthesis of porphyrin-based covalent organic polymers (PTZ-COP, CBZ-COP, and TPA-COP) through a one-pot aromatic electrophilic polycondensation reaction between pyrrole and monomeric aldehydes (PTZ-CHO, CBZ-CHO and TPA-CHO).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!