Background: To assess the efficacy of toric intraocular lenses (IOLs) in combined cataract and minimally invasive glaucoma surgery (MIGS), visual and refractive outcomes were compared between eyes implanted with non-toric and toric IOLs during iStent triple procedures.

Methods: In this retrospective study, open angle glaucoma eyes with preexisting corneal astigmatism of -1.5 diopter (D) or more and implanted with non-toric (n = 9) or toric (n = 9) IOLs were included. The main outcome measures were the intergroup difference in the uncorrected visual acuity (UCVA) and refractive astigmatism at 3 months postoperatively.

Results: Preoperatively, the logarithm of the minimum angle of resolution (logMAR) UCVAs and refractive astigmatism were equivalent between the groups. Postoperatively, the logMAR UCVA (non-toric, 0.45 ± 0.31; toric, 0.14 ± 0.15; P = 0.021) was significantly better and the refractive astigmatism (non-toric, -2.03 ± 0.63 D; toric, -0.67 ± 0.53 D; P = 0.0014) significantly less in the toric group. The toric group had postoperative improvements in the logMAR UCVA (-0.21, P = 0.020) and refractive astigmatism (+ 1.72 D, P = 0.0039). Vector analyses showed the postoperative centroid magnitude and confidence eclipses of refractive astigmatism was less in the toric group (0.47 D at 173°±0.73D) than the non-toric group (1.10 D at 2°±1.91D). Postoperatively, 78% of eyes in the toric group had 1.0 D or less refractive astigmatism compared with 11% in the non-toric group. Surgically induced astigmatism (non-toric, 0.18 D at 65°; toric, 0.29 D at 137°) did not differ between groups.

Conclusion: Use of toric IOLs is a reasonable option for better visual outcomes when the combined cataract and iStent surgery is performed in glaucomatous eyes with corneal astigmatism.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9749338PMC
http://dx.doi.org/10.1186/s12886-022-02707-1DOI Listing

Publication Analysis

Top Keywords

refractive astigmatism
24
toric group
16
toric
11
astigmatism
10
toric intraocular
8
visual acuity
8
glaucomatous eyes
8
combined cataract
8
implanted non-toric
8
toric iols
8

Similar Publications

Introduction: Implantable collamer lens (ICL) is a posterior chamber phakic intraocular lens. It is usually indicated for high refractive error correction that cannot be treated by laser vision correction due to physical limitations.

Case Presentation: 39 years old male with a past ocular history of keratoconus underwent astigmatic keratotomy followed by crosslinking in both eyes a couple of years later due to signs of corneal ectasia progression.

View Article and Find Full Text PDF

Purpose: To evaluate the clinical outcome of laser-assisted surgical correction of high hyperopic or mixed astigmatism using small incision intrastromal lenticule rotation (SMILERO) alone or combined with photorefractive keratectomy (PRK).

Methods: This retrospective case series enrolled 25 eyes with high astigmatism that underwent SMILERO surgery. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), uncorrected near visual acuity (UNVA), manifest refraction, central corneal thickness, and corneal higher order aberrations were analyzed before surgery and after 3, 6, and 12 months of follow-up.

View Article and Find Full Text PDF

Purpose: To evaluate a new regression-derived nomogram for high myopic astigmatism in small incision lenticule extraction (SMILE) surgery.

Methods: In this prospective study, data from 180 eyes with myopic astigmatism ranging from -2.50 to -4.

View Article and Find Full Text PDF

Purpose: To compare the clinical outcomes, surgical workflow, and patient satisfaction following small incision lenticule extraction (SMILE) performed with the VisuMax 800 in one eye and the VisuMax 500 in the contralateral eye (both Carl Zeiss Meditec).

Methods: This was a prospective, single-site clinical study of patients undergoing SMILE for myopia and myopic astigmatism between February 2022 and August 2023. Each patient underwent bilateral treatment using the VisuMax 800 (VM800 group) in one eye and the VisuMax 500 (VM500 group) in the contralateral eye.

View Article and Find Full Text PDF

Purpose: To evaluate residual refractive errors after intraocular lens (IOL) extraction and the safety and effectiveness of refractive correction procedures.

Methods: The eligibility criteria for this systematic review were patients who had undergone cataract or clear lens extraction and had experienced residual refractive error. All study designs were considered for inclusion and non-English publications, non-peer reviewed articles, books, and systematic reviews were excluded.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!