Background: Given the existence of multiple low-temperature thermoplastics, clinicians fabricating can readily modify an orthoses' thickness, weight and flexibility, among other properties. However, there is limited evidence on the impact of such different materials on upper extremities' biomechanics.
Objective: Our study aimed to investigate differences in joint stabilization and movement coordination provided by upper extremity orthotics fabricated with low-temperature thermoplastics of different thicknesses.
Study Design: Inferential, cross-sectional study.
Method: We conducted a kinematic analysis of a standardized task through a three-dimensional motion capture system. Ten participants (5 female) performed the same task under three circumstances: 1) wearing a volar wrist immobilization orthosis, made with a 3.2-mm thick low-temperature thermoplastic; 2) using the same orthotic fabricated with a 1.6-mm thick material; and (3) without orthoses. We divided the standardized task into five logical phases for data analysis, obtaining the active range of motion of the shoulder, elbow, forearm, and wrist joints as the primary outcome. Secondary outcomes included movement smoothness and coordination, measured by the number of motor units, time, and distance travelled by the upper extremity.
Results: Despite changes in thermoplastic thickness, both orthotics significantly restricted the wrist motion during task performance (F(2,16) = 14.32, P < .01, and η2p = 0.797), with no difference between the 2 devices and no significant changes to proximal joints' active range of motion. Although orthoses use increased the time required for task performance (F(2,16) = 23.05, P < .01, and η2p = 0.742), no significant differences in movement smoothness or coordination were noted.
Conclusion: Our results indicate that wrist orthoses fabricated with a 1.6-mm thick low-temperature thermoplastic can provide joint stabilization similar to a device made from a 3.2-mm thickness material, suggesting thinner thermoplastics' efficacy to stabilize joints in the absence of contractures or preexisting chronic conditions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/PXR.0000000000000162 | DOI Listing |
Neurosurg Rev
January 2025
Department of Biophysics, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, Zabrze, 41-808, Poland.
Atlantoaxial dislocation (AAD) is a serious condition in which the first two cervical vertebrae lose their anatomical position and stability. This may lead to neurological complications, including death. The treatment of AAD remains controversial, and posterior instrumentation with pedicle screw placement is one of the commonly used methods.
View Article and Find Full Text PDFFront Robot AI
January 2025
Life- and Neurosciences, Frankfurt Institute for Advanced Studies, Frankfurt am Main, Germany.
Biological vision systems simultaneously learn to efficiently encode their visual inputs and to control the movements of their eyes based on the visual input they sample. This autonomous joint learning of visual representations and actions has previously been modeled in the Active Efficient Coding (AEC) framework and implemented using traditional frame-based cameras. However, modern event-based cameras are inspired by the retina and offer advantages in terms of acquisition rate, dynamic range, and power consumption.
View Article and Find Full Text PDFRSC Adv
January 2025
National and Local Joint Engineering Research Center of Advanced Packaging Material Research and Development Technology, School of Packaging and Materials Engineering, Hunan University of Technology Zhuzhou 412007 Hunan China
To further improve the performance of PA66 and expand its applications, a new strategy was proposed to introduce an alicyclic structure into PA66 chain by the copolymerization method. Initially, 3,3'-dimethyl-4,4'-diaminodicyclohexylmethane (MACM) was reacted with 1,6-adipic acid to form MACM6 salt, and then, it was copolymerized with PA66 salt to synthesize PA66/MACM6 copolymers with alicyclic structures. PA66/MACM6 copolymers exhibited good thermal stabilities, and the presence of alicyclic structure had no significant effect on their thermal stabilities.
View Article and Find Full Text PDFPlant Methods
January 2025
Faculty of Agriculture, Agriculture and Forestry University, Bharatpur, 13712, Nepal.
Background: Crossover interactions stemming from phenotypic plasticity complicate selection decisions when evaluating hybrid maize with superior grain yield and consistent performance. Consequently, a two-year, region-wide investigation of 45 hybrids maize across Nepal was performed with the aim of disclosing both site and wide adapted hybrids. Utilizing an innovative "ProbBreed" package, based on Bayesian probability analysis of randomized complete block designs with three replicated trials at each station, this study substantively streamlines hybrids maize selection.
View Article and Find Full Text PDFCell Mol Biol Lett
January 2025
Jiangsu Key Laboratory of Neuropsychiatric Diseases and College of Pharmaceutical Sciences, Jiangsu Province Engineering Research Center of Precision Diagnostics and Therapeutics Development, Jiangsu Key Laboratory of Preventive and Translational Medicine for Geriatric Diseases, Suzhou Key Laboratory of Drug Research for Prevention and Treatment of Hyperlipidemic Diseases, Soochow University, 199 Ren'ai Road, Suzhou, 215123, Jiangsu, China.
Background: The protein cereblon (CRBN) mediates the antileukemia effect of lenalidomide (Len). Len binds to CRBN, recruits IKZF1/IKZF3, and promotes their ubiquitination and degradation, through which Len exhibits its antileukemia and antimyeloma activity. Therefore, the protein level of CRBN might affect the antiproliferative effect of Len.