Feasibility and safety of a new dedicated biliary stent for EUS-guided hepaticogastrostomy: The FIT study (with video).

Endosc Ultrasound

Department of Gastroenterology, Humanitas Clinical and Research Center, Digestive Endoscopy Unit, IRCCS, Rozzano; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy.

Published: January 2023

Background And Objectives: EUS-guided biliary drainage (BD) through hepaticogastrostomy (HGS) is an option in case of ERCP failure. Available data suggest that this procedure may be challenging with possible severe adverse events (AEs) mainly due to stent migration. The aim of our pilot study was to prospectively assess the technical and clinical outcomes of EUS-HGS using a new dedicated partially covered self-expandable metal stent with anti-migratory systems.

Methods: This is a single-center prospective study enrolling patients with malignant biliary obstruction undergoing EUS-HGS after failed ERCP, between June 2020 and March 2021. The primary endpoint was the technical success rate. Evaluation of specific stent-related technical features as compared with commonly used self-expandable metal stent, clinical success rate, and procedure-related AEs was also assessed.

Results: Twenty-two patients (15%-68.2%, female; mean age, 66.0 ± 10.0) were enrolled in the study analysis. Different causes of ERCP failure were infiltration of papilla by neoplastic tissue (4, 18.2%), unreachable papilla for duodenal stricture (9, 40.9%), surgically altered anatomy with Roux-en-Y reconstruction (4, 18.2%), and incomplete BD after transpapillary stent placement (5, 22.7%). Technical success was achieved in all patients, in a mean procedural time of 43.3 ± 26.8 min. Technical features were graded as high or medium in all cases. The clinical success rate was 91% (20/22, mean follow-up: 10.8 ± 3.1 months). There were no cases of stent misplacement or stent migration. Three (13.6%) cases of a hepatic abscess requiring percutaneous drainage and systemic antibiotics were reported, with no impact on clinical success and following oncologic treatments. No deaths occurred.

Conclusion: EUS-HGS with a new dedicated stent with anti-migratory systems is feasible and effective, preventing stent migration, and misplacement. Although the persistent procedural challenges, dedicated devices may contribute to outcomes improvement and procedure diffusion.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10134938PMC
http://dx.doi.org/10.4103/EUS-D-22-00023DOI Listing

Publication Analysis

Top Keywords

stent migration
12
success rate
12
clinical success
12
stent
9
ercp failure
8
eus-hgs dedicated
8
self-expandable metal
8
metal stent
8
stent anti-migratory
8
technical success
8

Similar Publications

Introduction: Bowel perforation due to migrated biliary stent is a rare complication. Here, we report a case of duodenal and ascending colonal perforation due to biliary stent migration.

Case Presentation: A 35-year-old man is complaining of right upper abdominal pain presented to the gastroenterology department.

View Article and Find Full Text PDF

Stanford type B aortic dissection involving the left subclavian artery (LSA) poses significant clinical challenges. The Castor single-branch stent graft and in situ fenestration are commonly used techniques, but the better endovascular treatment remains debated. This study evaluates the clinical effects of the Castor single-branched stent graft versus in situ fenestration in treating Stanford type B aortic dissection involving the LSA.

View Article and Find Full Text PDF

Multi-disciplinary treatment of broncho-esophageal fistula in a high-risk single-lung patient.

J Cardiothorac Surg

January 2025

Section of Cardiothoracic Surgery, Department of Heart Disease, Haukeland University Hospital, Jonas Lies vei 65, 5021, Bergen, Norway.

Background: A broncho-esophageal fistula (BEF) is a medical and surgical disaster. Treatment of BEF is often limited to palliative stent treatment that may migrate or cause erosions and tissue necrosis. Surgical repair of BEF is the only established definite treatment.

View Article and Find Full Text PDF

Purpose: To assess whether safety profile and treatment success of percutaneous biodegradable biliary stent placement are competitive with traditional treatment options for treatment of benign biliary strictures.

Materials And Methods: PubMed and EMBASE were systematically reviewed for articles reporting percutaneous biodegradable stent placement for treating benign biliary strictures. Databases were searched for articles until December 2023, with the earliest included article dating from April 2016.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!