Context: Pulse oximetry is one of the most revolutionary methods used to monitor the patients in the clinical setting, particularly intensive care and anesthesia. We usually use the index or middle finger to measure SpO values, but little is known about the inter-digital differences that exist between the fingers.

Aim: We aim to compare the peripheral capillary oxygen saturation among fingers of both hands. A total of 93 healthy volunteers aged between 18 and 50 years participated in the study.

Materials And Methods: Their SpO values were recorded from each finger with at least 1 minute of resting interval. Their blood pressure, heart rate, and body temperature were recorded as well.

Result: A total of 930 measurements were obtained from 93 volunteers. The highest average SpO value of right-handed volunteers was measured from the left little finger (98.48 ± 0.62) of right-handed volunteers, and it was statistically significant when compared with the right ring finger, right little finger, left thumb, left index, left middle finger, and left ring finger. The highest average SpO from left-handed volunteers was obtained from the right index finger, but it was statistically insignificant.

Conclusion: We assume ethnic and climatic differences to play a role in contradictory results noted from previous studies conducted, and this needs to be investigated further. It is recommended that multiple readings may be obtained from other fingers as well before coming to any conclusion as inter-finger variability cannot be ignored.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9730959PMC
http://dx.doi.org/10.4103/jfmpc.jfmpc_2596_20DOI Listing

Publication Analysis

Top Keywords

finger
8
middle finger
8
spo values
8
highest average
8
average spo
8
right-handed volunteers
8
ring finger
8
finger left
8
volunteers
5
left
5

Similar Publications

Flexible smart sensing materials are gaining tremendous momentum in wearable and bionic smart electronics. To satisfy the growing demand for sustainability and eco-friendliness, biomass-based hydrogel sensors for green and biologically safe wearable sensors have attracted significant attention. In this work, we have prepared MCC/PAA/AgNWs/CNTs hydrogel sensors with excellent conductive sensing properties by a simple physical blending method.

View Article and Find Full Text PDF

Retrograde intramedullary headless compression screw fixation for pediatric mid-diaphyseal proximal phalanx malunion: A case study.

Int J Surg Case Rep

January 2025

Department of Orthopaedics and Traumatology, AZ Delta, Deltalaan 1, 8800 Roeselare, Belgium; Department of Cardio and Organ Systems, Hasselt University, Martelarenlaan 42, 3500 Hasselt, Belgium. Electronic address:

Introduction: Proximal phalanx fractures in children, especially mid-diaphyseal fractures, can result in malunion and significant functional impairment. Early malunions require prompt and effective intervention to prevent long-term complications. This case study highlights the use of intramedullary headless compression screw (IMHCS) fixation in addressing a proximal phalanx malunion.

View Article and Find Full Text PDF

Fatal systemic bacterial infections in two harbour porpoises (Phocoena phocoena) secondary to bite wounds from grey seals (Halichoerus grypus).

J Comp Pathol

January 2025

Scottish Marine Animal Stranding Scheme, School of Biodiversity, One Health and Veterinary Medicine, College of Medical, Veterinary and Life Sciences, University of Glasgow, Glasgow G12 8QQ, Scotland, UK.

Mycoplasma phocicerebrale, the causative agent of seal ('speck') finger, a zoonotic disease, is a common commensal in the oral cavity of various seal species. Historically associated with seal hunters, it remains a significant risk for those handling or rehabilitating marine mammals. While primarily known for causing severe cellulitis in humans, M.

View Article and Find Full Text PDF

Background And Purpose: To develop a new test to assess the motor coordination of the upper limbs, and to investigate the test-retest and inter-rater reliability, construct validity, standard error of measurement (SEM), minimum detectable change (MDC), and the reference values.

Methods: The Upper Extremity Motor Coordination Test (UEMOCOT)was applied for 20 s, with the individual touching two targets (one right and one left) as quickly as possible, first with the hand (manual task) and then with the index finger (finger task). To test-retest reliability, the UEMOCOT was administered two times.

View Article and Find Full Text PDF

Bio-inspired carbon-based artificial muscle with precise and continuous morphing capabilities.

Natl Sci Rev

January 2025

CAS Key Laboratory of Organic Solids, Beijing National Laboratory for Molecular Sciences (BNLMS), CAS Research/Education Center for Excellence in Molecular Sciences, Institute of Chemistry, Chinese Academy of Sciences, Beijing 100190, China.

In the face of advancements in microrobotics, intelligent control and precision medicine, artificial muscle actuation systems must meet demands for precise control, high stability, environmental adaptability and high integration miniaturization. Carbon materials, being lightweight, strong and highly conductive and flexible, show great potential for artificial muscles. Inspired by the butterfly's proboscis, we have developed a carbon-based artificial muscle, hydrogen-substituted graphdiyne muscle (HsGDY-M), fabricated efficiently using an emerging hydrogen-substituted graphdiyne (HsGDY) film with an asymmetrical surface structure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!