AI Article Synopsis

  • Radiation therapy is advancing towards more precise dose delivery, and combining online MR imaging with particle therapy (like protons or carbon ions) could enhance this precision despite challenges in treatment planning and workflow.
  • The aim of this study was to develop a workflow for commissioning a light ion beam line that integrates a dipole magnet for MR-guided particle therapy (MRgPT), enabling accurate dose calculations using high-quality magnetic field maps.
  • The research conducted involved measuring beam data and developing a 3D model for magnetic fields, leading to findings that a specific magnetic field volume significantly impacts beam trajectory and particle range accuracy for protons and carbon ions.

Article Abstract

Background: Radiation therapy is continuously moving towards more precise dose delivery. The combination of online MR imaging and particle therapy, for example, radiation therapy using protons or carbon ions, could enable the next level of precision in radiotherapy. In particle therapy, research towards a combination of MR and particle therapy is well underway, but still far from clinical systems. The combination of high magnetic fields with particle therapy delivery poses several challenges for treatment planning, treatment workflow, dose delivery, and dosimetry.

Purpose: To present a workflow for commissioning of a light ion beam line with an integrated dipole magnet to perform MR-guided particle therapy (MRgPT) research, producing not only basic beam data but also magnetic field maps for accurate dose calculation. Accurate dose calculation in magnetic field environments requires high-quality magnetic field maps to compensate for magnetic-field-dependent trajectory changes and dose perturbations.

Methods: The research beam line at MedAustron was coupled with a resistive dipole magnet positioned at the isocenter. Beam data were measured for proton and carbon ions with and without an applied magnetic field of 1 T. Laterally integrated depth-dose curves (IDC) as well as beam profiles were measured in water while beam trajectories were measured in air. Based on manufacturer data, an in silico model of the magnet was created, allowing to extract high-quality 3D magnetic field data. An existing GATE/Geant4 Monte Carlo (MC) model of the beam line was extended with the generated magnetic field data and benchmarked against experimental data.

Results: A 3D magnetic field volume covering fringe fields until 50 mT was found to be sufficient for an accurate beam trajectory modeling. The effect on particle range retraction was found to be 2.3 and 0.3 mm for protons and carbon ions, respectively. Measured lateral beam offsets in water agreed within 0.4 and -0.5 mm with MC simulations for protons and carbon ions, respectively. Experimentally determined in-air beam trajectories agreed within 0.4 mm in the homogeneous magnetic field area.

Conclusion: The presented approach based on in silico modeling and measurements allows to commission a beam line for MRgPT while providing benchmarking data for the magnetic field modeling, required for state-of-the art dose calculation methods.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.16143DOI Listing

Publication Analysis

Top Keywords

magnetic field
32
particle therapy
24
carbon ions
16
protons carbon
12
beam
11
magnetic
10
mr-guided particle
8
therapy
8
radiation therapy
8
dose delivery
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!