Validation of three-dimensional printed models of intracranial aneurysms.

Interv Neuroradiol

Neuroradiology, Hôpital Güi-de-Chauliac, CHU de Montpellier, Montpellier, France.

Published: October 2024

Introduction: Three-dimensional (3D) printing has evolved for medical applications as it can produce customized 3D models of devices and implants that can improve patient care. In this study, we aimed to validate the geometrical accuracy of the 3D models of intracranial aneurysms printed using Stereolithography 3D printing technology.

Materials And Methods: To compare the unruptured intracranial aneurysm mesh between the five patients and 3D printed models, we opened the DICOM files in the Sim&Size® simulation software, selected the region of interest, and performed the threshold check. We juxtaposed the 3D reconstructions and manually rotated the images to get the same orientation when needed and measured deviations at different nodes of the patient and 3D printed model meshes.

Results: In the first patient, 80% of the nodes were separated by <0.56 mm and 0.17 mm. In the second patient, the deviations were below 0.17 mm for 80% of the meshes' nodes. In the next three patients, the deviations were below 0.21, 0.23, and 0.11 mm for 80% of the meshes' nodes. Finally, the overall deviation was below 0.21 mm for 80% of the mesh nodes of the five aneurysms.

Conclusions: 3D printed models of intracranial aneurysms are accurate, having surfaces that resemble that of patients' angiographies with an 80% cumulative deviation below 0.21 mm.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11569489PMC
http://dx.doi.org/10.1177/15910199221143254DOI Listing

Publication Analysis

Top Keywords

printed models
8
models intracranial
8
intracranial aneurysms
8
validation three-dimensional
4
printed
4
three-dimensional printed
4
models
4
aneurysms introduction
4
introduction three-dimensional
4
three-dimensional printing
4

Similar Publications

Background: Accurate femoral tunnel positioning is essential for successful anterior cruciate ligament (ACL) reconstruction. Tunnel malposition can happen due to limited arthroscopic visibility as well as anatomic variance. The use of customized patient-specific guides can optimize surgical planning and enhance accuracy.

View Article and Find Full Text PDF

This study investigated the potential pulmonary toxicity of polycarbonate (PC) emissions from fused filament fabrication (FFF) three-dimensional printing (3DP) via inhalation in Sprague Dawley rats. Previously, our results demonstrated no significant pulmonary effects following exposure to a 0.5 mg/m PC.

View Article and Find Full Text PDF

Clinical practice guidelines (CPGs) are shared through various dissemination strategies using a range of dissemination products and channels. However, users may have different needs for accessing and understanding them. Patients and carers from low- and middle-income countries might face challenges in accessing CPGs such as inadequate systems for printed book distribution and insufficient and substandard photocopies.

View Article and Find Full Text PDF

Background: Dental education aims to equip students with the knowledge and skills required to competently manage clinical scenarios. It is critical that dental education evolves with technological advances to ensure quality student preparation. New technologies such as three-dimensional printing (3DP) have found their way into dental education and could overcome the limitations of traditional methods.

View Article and Find Full Text PDF

Analysis of midsole gait in running shoes with various 3d printed biomimetic structure.

Sci Rep

March 2025

Department Fashion and Textiles, Dong-A University, Busan, 49315, Republic of Korea.

This study aims to analyze the impact of different 3D printed biomimetic midsole structures on gait phase and gait ratio to optimize running shoe design. Participants wore shoes with various 3D printed midsoles while gait ratio(heel rotation, foot balance, forefoot balance, hallux stiffness, medial forefoot balance, and meta loading) and gait phase(forefoot off phase, foot flat phase, forefoot contact phase, and initial contact phase) data were collected using a gait analysis system. These metrics reflect foot pressure distribution, gait symmetry, and balance, providing insights into how midsole structures affect gait.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!