This study assessed 3 endodontic motors, X-Smart Plus (Dentsply Sirona, Ballaigues, Switzerland), VDW.Silver Reciproc (VDW GmbH, München, Germany) and, iRoot (Bassi Endodontics, Belo Horizonte, Brazil) in 2 different reciprocating settings. The movements evaluated were 170° in counter-clockwise (CCW) and 50° in clockwise (CW) at 350 RPM, and 150° CCW and 30° CW at 300 RPM. For the X-Smart Plus and VDW Silver the settings used were the ones in the motor library. For the iRoot, the motor was adjusted to the angles of the study. A customized optic target was attached to the contra-angle of the motor and the movements were recorded with a high-resolution camera (K2 DistaMaxTM Long-Distance Microscope System, Infinity Photo-Optical Company, Colorado, EUA) at 2,400 frames per second (FPS). The images were analyzed with the Vision Research software (Inc. Headquarters, Wayne, New Jersey, EUA). The following kinematic parameters were assessed: CCW angle, CW angle, speed (RPM) at both directions, and, standstill time at each change of directions. The Intraclass Correlation Coefficient (ICC) and Kruskal-Wallis (method of Dunn) were used at a significant level of 5%. There was no statistically significant difference among the motors at the 150°/30° setting (P > .05); the iRoot was the least reliable at the 170°/50° setting for CCW angle, speed, and net angle parameters (P < 0.05). The standstill time of all motors in both directions was identical. None of the motors were able to reproduce faithfully the set movements. The iRoot motor presented a higher discrepancy when compared to X-Smart and VDW Silver.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9733375PMC
http://dx.doi.org/10.1590/0103-6440202204855DOI Listing

Publication Analysis

Top Keywords

x-smart vdw
12
vdw silver
12
endodontic motors
8
iroot motor
8
ccw angle
8
angle speed
8
standstill time
8
iroot
5
motors
5
reciprocating kinematics
4

Similar Publications

The present study evaluated the actual rotational speed of three different endodontic motors compared to the values provided by the manufacturers. A total of three endodontic motors (X-Smart Plus, VDW.Silver, and iRoot) were tested at 400 rpm and 800 rpm and 2 N/cm2 torque.

View Article and Find Full Text PDF

This study assessed 3 endodontic motors, X-Smart Plus (Dentsply Sirona, Ballaigues, Switzerland), VDW.Silver Reciproc (VDW GmbH, München, Germany) and, iRoot (Bassi Endodontics, Belo Horizonte, Brazil) in 2 different reciprocating settings. The movements evaluated were 170° in counter-clockwise (CCW) and 50° in clockwise (CW) at 350 RPM, and 150° CCW and 30° CW at 300 RPM.

View Article and Find Full Text PDF

Impact of a modified motion on the fatigue life of NiTi reciprocating instruments: a Weibull analysis.

Clin Oral Investig

July 2019

School of Dentistry, Endodontic Clinical Section, Master in Clinical Endodontology, Department of Biomedical and Neuromotor Sciences(DIBINEM), University of Bologna, Alma Mater Studiorum, Bologna, Italy.

Objectives: To evaluate the impact of a modified motion on the fatigue life of four brands of nickel-titanium (NiTi) reciprocating instruments.

Materials And Methods: Cyclic fatigue (CF) resistance of 160 instruments was evaluated in an artificial stainless-steel canal (90° angle, 5-mm radius of curvature). WaveOne and WaveOne Gold (Denstply Maillefer, Baillagues, Switzerland) and Reciproc and Reciproc Blue (VDW, Munich, Germany) were tested with two different motions: (1) X-Smart Plus (Denstply Maillefer) endodontic motor and (2) a 4:1 contra-angle with an experimental motion (EVO) with different rotation angles and based on a sinusoidal acceleration.

View Article and Find Full Text PDF

Introduction: The aim of the present study was to evaluate the cyclic fatigue resistance of the WaveOne Gold Glider (Dentsply Maillefer, Ballaigues, Switzerland), R-Pilot (VDW, Munich, Germany), and ProGlider (Dentsply Maillefer) glide path instruments.

Methods: Ten instruments from each glide path system were included in the study. A stainless steel artificial canal with a 90° angle and a 3-mm radius of curvature was used for the testing process.