Background And Purpose: The ThrombX Retriever is a novel mechanical thrombectomy device that adjusts the distance between two mesh segments to axially hold thrombus. A post-market study assessed safety and performance in acute ischemic stroke patients with large artery occlusion.
Methods: A single-arm prospective multi-center study enrolled patients at 5 European Centers. Patients had a symptomatic large-artery occlusion of the intracranial Internal Carotid or the Middle Cerebral Artery, M1 segment. The primary outcome measure was the modified treatment in cerebral infarction (mTICI) score, on the immediate post-procedure angiogram after up to three device passes. Key secondary outcome measures were the mTICI score after a single pass and functional independence, defined as an mRS score ≤ 2 at 90 days.
Results: Thirty patients (16 Females, mean age 72 years), with NIHSS 4-25 (mean 15.5) were treated. Twenty-eight (93%) achieved mTICI 2b-3 within 3 passes, and 24 (80%) were with the first pass (FP). FP mTICI 2c-3 reperfusion was achieved in 19 (63%) cases. Seventeen of 24 (71%) patients treated with a balloon guide and the ThrombX Retriever had a FP mTICI 2c-3 reperfusion. After all interventions, mTICI 2b-3 was seen in 30 (100%) patients. Twenty-one of the 29 (73%) patients with 90-day follow-up were functionally independent (mRS≤2). No device-related serious adverse events were observed.
Conclusion: This preliminary study suggests the ThrombX Retriever is safe and has a high rate of substantial reperfusion. A larger prospective trial to assess the device effectiveness is planned.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11095356 | PMC |
http://dx.doi.org/10.1177/15910199221118146 | DOI Listing |
J Neurointerv Surg
April 2024
Neurology and Radiology, Emory University School of Medicine, Atlanta, Georgia, USA.
Background: Mechanical thrombectomy (MT) has become standard for large vessel occlusions, but rates of complete recanalization are suboptimal. Previous reports correlated radiographic signs with clot composition and a better response to specific techniques. Therefore, understanding clot composition may allow improved outcomes.
View Article and Find Full Text PDFInterv Neuroradiol
April 2024
Department of Neurointerventions, National Institute of Mental Health, Neurology and Neurosurgery, Budapest, Hungary.
Background And Purpose: The ThrombX Retriever is a novel mechanical thrombectomy device that adjusts the distance between two mesh segments to axially hold thrombus. A post-market study assessed safety and performance in acute ischemic stroke patients with large artery occlusion.
Methods: A single-arm prospective multi-center study enrolled patients at 5 European Centers.
Cardiovasc Intervent Radiol
February 2021
Department of Radiology, New England Center for Stroke Research, University of Massachusetts Medical School, 55 Lake Ave N, SA-107R, Worcester, MA, 01655, USA.
Purpose: The first-pass effect during mechanical thrombectomy improves clinical outcomes regardless of first-line treatment approach, but current success rates for complete clot capture with one attempt are still less than 40%. We hypothesize that the ThrombX retriever (ThrombX Medical Inc.) can better engage challenging clot models during retrieval throughout tortuous vasculature in comparison with a standard stent retriever without increasing distal emboli.
View Article and Find Full Text PDFJ Neurointerv Surg
September 2021
Neurosurgery, Prisma Health Upstate, Greenville, South Carolina, USA.
Background: Direct aspiration thrombectomy techniques use large bore aspiration catheters for mechanical thrombectomy. Several aspiration catheters are now available. We report a bench top exploration of a novel beveled tip catheter and our experience in treating large vessel occlusions (LVOs) using next-generation aspiration catheters.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!