Quantifying 6D tumor motion and calculating PTV margins during liver stereotactic radiotherapy with fiducial tracking.

Front Oncol

National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital and Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen, China.

Published: November 2022

AI Article Synopsis

Article Abstract

Objective: Our study aims to estimate intra-fraction six-dimensional (6D) tumor motion with rotational correction and the related correlations between motions of different degrees of freedom (DoF), as well as quantify sufficient anisotropic clinical target volume (CTV) to planning target volume (PTV) margins during stereotactic body radiotherapy (SBRT) of liver cancer with fiducial tracking technique.

Methods: A cohort of 12 patients who were implanted with 3 or 4 golden markers were included in this study, and 495 orthogonal kilovoltage (kV) pairs of images acquired during the first fraction were used to extract the spacial position of each golden marker. Translational and rotational motions of tumor were calculated based on the marker coordinates by using an iterative closest point (ICP) algorithm. Moreover, the Pearson product-moment correlation coefficients () were applied to quantify the correlations between motions with different degrees of freedom (DoFs). The population mean displacement ( ), systematic error (Σ) and random error () were obtained to calculate PTV margins based on published recipes.

Results: The mean translational variability of tumors were 0.56, 1.24 and 3.38 mm in the left-right (LR, ), anterior-posterior (AP, ), and superior-inferior (SI, ) directions, respectively. The average rotational angles , and around the three coordinate axes were 0.88, 1.24 and 1.12, respectively. (||>0.4) was obtainted between - , - , - and - . The PTV margins calculated based on 13 published recipes in X, Y, and Z directions were 1.08, 2.26 and 5.42 mm, and the 95% confidence interval (CI) of them were (0.88,1.28), (1.99,2.53) and (4.78,6.05), respectively.

Conclusions: The maximum translational motion was in SI direction, and the largest correlation coefficient of - was obtained. We recommend margins of 2, 3 and 7 mm in LR, AP and SI directions, respectively.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9709257PMC
http://dx.doi.org/10.3389/fonc.2022.1021119DOI Listing

Publication Analysis

Top Keywords

ptv margins
16
tumor motion
8
fiducial tracking
8
correlations motions
8
motions degrees
8
degrees freedom
8
target volume
8
calculated based
8
based published
8
margins
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!