Background: Ultrasound imaging is used for diagnosis, treatment, and blood vessel visualization during venous catheter placement. However, various physiological factors (e.g., body temperature and exercise) influence vein diameters, which are expected to exhibit daily or diurnal fluctuations. Therefore, this study aimed to determine the intraday (short-term) and interday (long-term) reproducibility of repeated measurements of the depth and diameter of peripheral superficial veins.

Methods: Twenty-three healthy young women (mean age, 21.7 ± 0.8 years) participated in the study to examine the short- and long-term reproducibility of the depth and diameter of the cutaneous vein in the left elbow fossa acquired by ultrasound imaging. Short-term measurement intervals were 10 s, and the probe was released from the skin for each acquisition, which was repeated five consecutive times. Long-term measurements were performed at the same time on the next day following the same procedure. The acquired images were analyzed for vein depth and diameter using ImageJ software. The intraclass correlation coefficient (ICC) was calculated to determine the short- and long-term reproducibility of the measurements. The relationship between the venous depth and venous diameter intra-individual variation was analyzed, as well as the influence of body composition (body fat and muscle mass) on the venous diameter and depth.

Results: For vein depth measurements, the short- and long-term ICCs were 0.94-0.96 and 0.88, respectively. For the vein diameter, the short- and long-term ICCs were 0.94-0.97 and 0.67, respectively. The short-term ICCs for both vein depth and diameter exceeded 0.9, indicating that the ultrasound vascular measurement was sufficiently reliable. However, long-term reproducibility was slightly lower, especially for the vein diameter. No correlation was found between the intra-individual variation of the vein diameter and vein depth. Although the vein diameter and body fat mass uncorrelated, the vein depth and body fat mass significantly correlated (r = 0.675, 95% confidence interval = 0.281-0.830).

Conclusions: The long-term reproducibility of vein diameters was somewhat lower than that of the short-term reproducibility. This could be attributed to fluctuations in the physiological state of the participant rather than to the instability of the measurement. Therefore, ultrasound measurement of the peripheral superficial vein is sufficiently reliable.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9719193PMC
http://dx.doi.org/10.1186/s12880-022-00945-9DOI Listing

Publication Analysis

Top Keywords

long-term reproducibility
24
vein depth
24
short- long-term
20
depth diameter
20
vein diameter
16
vein
14
peripheral superficial
12
ultrasound imaging
12
body fat
12
diameter
11

Similar Publications

A reliable and objective method of measuring soft tissue changes in 2D photographs after distraction osteogenesis in individuals with CL/P.

J Plast Surg Hand Surg

January 2025

Department of Clinical Sciences in Malmö, Lund University, Lund, Sweden; Department of Oral and Maxillofacial Surgery, Skåne University Hospital, Malmö, Sweden; Department of Orthodontics, Malmö University, Malmö, Sweden.

Pre- and postoperative photos to assess results are widely used in plastic and reconstructive surgery, for instance, in patients with cleft lip and palate (CL/P). Evaluations are often performed by assessment panels by viewing the photos. However, these are prone to be subjective.

View Article and Find Full Text PDF

Objective: Aortic valve repair/sparing have been established as effective treatments for aortic regurgitation and/or aortic aneurysms. However, concerns remain regarding long-term durability, reproducibility, and patient selection. This study aims to asses long-term clinical and echocardiographic outcomes, with a focus on aortic regurgitation grade and left ventricular ejection fraction evolution, in adults undergoing these procedures.

View Article and Find Full Text PDF

Integrated spaceflight transcriptomic analyses and simulated space experiments reveal key molecular features and functional changes driven by space stressors in space-flown C. elegans.

Life Sci Space Res (Amst)

February 2025

Institute of Environmental Systems Biology, College of Environmental Science and Engineering, Dalian Maritime University, Dalian, 116026, Liaoning, PR China.

The space environment presents unique stressors, such as microgravity and space radiation, which can induce molecular and physiological changes in living organisms. To identify key reproducible transcriptomic features and explore potential biological roles in space-flown C. elegans, we integrated transcriptomic data from C.

View Article and Find Full Text PDF

Deep eutectic solvent induced silver-gel as a flexible SERS substrate for sensitive detection of antibiotics under low temperature conditions.

J Hazard Mater

January 2025

State Key Laboratory of Chemistry and Utilization of Carbon Based Energy Resources; College of Chemistry, Xinjiang University, Urumqi, 830017, Xinjiang, PR China; School of Materials Science and Engineering, Xinjiang Engineering Research Center of Environmental and Functional Materials, Xinjiang University, Urumqi, 830017, Xinjiang, PR China. Electronic address:

Antibiotic residues pose a significant threat to global health. Traditional detection methods for antibiotics are cumbersome, time-consuming and often incapable of achieving non-destructive detection at low temperatures. This research introduces a groundbreaking innovation in antibiotic detection: a flexible Surface-Enhanced Raman Scattering substrate based on a silver composite deep eutectic solvent (DES) gel, specifically engineered for low-temperature antibiotic detection.

View Article and Find Full Text PDF

Background: Understanding fetal growth is essential for predicting perinatal outcomes and long-term health implications. This study explores the correlation between ultrasound parameters and gestational age in the Minangkabau ethnic group, focusing on the biometric variables biparietal diameter (BPD), head circumference (HC), abdominal circumference (AC), femur length (FL), and humerus length (HL) during the second trimester.

Methods: Conducted from December 2020 to December 2021, the research employed a cross-sectional design at the Fetomaternal clinic of Hospital M.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!