Background: In external beam radiotherapy, a prediction model is required to compensate for the temporal system latency that affects the accuracy of radiation dose delivery. This study focused on a thorough comparison of seven deep artificial neural networks to propose an accurate and reliable prediction model.
Methods: Seven deep predictor models are trained and tested with 800 breathing signals. In this regard, a nonsequential-correlated hyperparameter optimization algorithm is developed to find the best configuration of parameters for all models. The root mean square error (RMSE), mean absolute error, normalized RMSE, and statistical F-test are also used to evaluate network performance.
Results: Overall, tuning the hyperparameters results in a 25%-30% improvement for all models compared to previous studies. The comparison between all models also shows that the gated recurrent unit (GRU) with RMSE = 0.108 ± 0.068 mm predicts respiratory signals with higher accuracy and better performance.
Conclusion: Overall, tuning the hyperparameters in the GRU model demonstrates a better result than the hybrid predictor model used in the CyberKnife VSI system to compensate for the 115 ms system latency. Additionally, it is demonstrated that the tuned parameters have a significant impact on the prediction accuracy of each model.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10018664 | PMC |
http://dx.doi.org/10.1002/acm2.13854 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!