Purpose: We have developed and validated a high-sensitivity method to quantify lysergic acid diethylamide (LSD) and 2-oxo-3-hydroxy-LSD (OH-LSD) in oral fluid samples using liquid-liquid extraction and liquid chromatography-tandem mass spectrometry (LC‒MS/MS). The method was applied to the quantification of both substances in 42 authentic oral fluid samples.

Methods: A liquid-liquid extraction was performed using 500 µL each of samples (oral fluid samples collected using Quantisal™ device) and dichloromethane/isopropanol mixture (1:1, v/v). Enzymatic hydrolysis was evaluated to cleave glucuronide metabolites.

Results: The limit of quantification was 0.01 ng/mL for both LSD and OH-LSD. The linearity was assessed between 0.01 and 5 ng/mL. Imprecision and bias were not higher than 10.2% for both analytes. Extraction recovery was higher than 69%. The analytes were stable in the autosampler at 10 °C for 24 h, and up to 30 days at 4 and -20 °C. The method was applied to the analysis of 42 oral fluid samples. LSD was detected in all samples (concentrations between 0.02 and 175 ng/mL), and OH-LSD was detected in 20 samples (concentrations between 0.01 and 1.53 ng/mL).

Conclusions: A high-sensitive method was fully validated and applied to authentic samples. To our knowledge, this is the first work to report concentrations of LSD and OH-LSD in authentic oral fluid samples.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11419-022-00622-0DOI Listing

Publication Analysis

Top Keywords

oral fluid
24
fluid samples
16
high-sensitivity method
8
lsd 2-oxo-3-hydroxy-lsd
8
mass spectrometry
8
samples
8
liquid-liquid extraction
8
method applied
8
authentic oral
8
lsd oh-lsd
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!