AI Article Synopsis

  • The study aimed to evaluate a new cutting, progressive thread design for implants to improve stability in low-density bone.
  • Four types of dental implants with different designs were tested in synthetic bone to measure their insertion torque, stability, and time taken for placement.
  • Results indicated that the PrimeTaper implant's design achieved similar primary stability in low-density bone as some aggressive designs while allowing for faster insertion, suggesting it could be a viable option for implant placement in challenging conditions.

Article Abstract

Purpose: To evaluate an alternative cutting, progressive thread design to increase primary stability in low-density bone.

Materials And Methods: Four different implants with different macrogeometries (Nobel Biocare Active [NA] Internal RP implants [4.3 × 11.5 mm], Straumann BLX Roxolid RB implants [4.5 × 10 mm], Astra Tech Implant EV implants [4.2 × 11 mm], and PrimeTaper [PT 4.2 × 11 mm]) were placed in simulated osteotomies and extraction sockets in synthetic bone (Sawbones) according to the manufacturers' protocol. Insertion torque and ISQ values were measured using Implantmed Plus motor and Ostell IDX, respectively. Insertion time was recorded. Average values were calculated and compared using ANOVA and Tukey test.

Results: Insertion torque (range: 5 to 44 Ncm) increased with increasing synthetic bone density for all implants. Different ISQ values in synthetic low-density bone were not observed in higher-density synthetic bone. Insertion torque of all implants was reduced when implants were placed in simulated sockets compared to simulated osteotomies. In both low-density and higher-density synthetic bone, the primary stability of PrimeTaper implants with cutting and progressive thread design was equivalent to that of the Nobel Biocare NobelActive implant with compressive thread design and greater than the BLX implant with compressive thread design.

Conclusion: Different implant macrogeometries obtain relatively high primary stability in low-density bone when measured by ISQ. Doublethread implant designs reduce insertion times in higher-density bone. A cutting and progressive compressing thread design provides density-sensing performance compared to aggressive condensing thread designs. This macrogeometry can achieve high primary stability associated with modest insertion torque compared to aggressive threaded implant designs known to attain the highest insertion torque. The presence of multiple cutting threads may offer advantages in obtaining primary stability in low-density bone.

Download full-text PDF

Source
http://dx.doi.org/10.11607/jomi.9656DOI Listing

Publication Analysis

Top Keywords

primary stability
24
insertion torque
20
thread design
16
synthetic bone
16
cutting progressive
12
stability low-density
12
low-density bone
12
progressive thread
8
implants
8
nobel biocare
8