Human leukocyte antigen-B27 typing by flow cytometry: Comparison of three CE-IVD methods.

Cytometry B Clin Cytom

Laboratoire d'immunologie clinique, Hôpital Nord, Saint-Étienne, France.

Published: November 2023

AI Article Synopsis

  • The study compares the effectiveness of three flow cytometry kits for diagnosing human leukocyte antigen B27 (HLA-B27) against traditional molecular biology methods.
  • Results showed that while all kits performed well, adjustments were necessary for accuracy, with IOTest™ and DuraClone B27™ exhibiting better outcomes than BD HLA-B27™.
  • Ultimately, DuraClone B27™ was deemed the most reliable, though genotyping remains crucial for cases where flow cytometry results are inconclusive.

Article Abstract

Background: The human leukocyte antigen B27 (HLA-B27), associated with chronic inflammatory diseases is thus widely performed for diagnostic purposes. Genotyping by molecular biology is the current gold standard for HLA-B27 typing, but cheaper and faster flow cytometry methods have been developed.

Methods: In this report, we compare analytical performances of three CE-IVD flow cytometry kits: IOTest™ and DuraClone B27™ from Beckman Coulter and BD HLA-B27™ from BD Biosciences on a Navios™ cytometer as compared to molecular biology.

Results: Routine analyses could conclude for HLA-B27 in197 patients (23.2%) and was not conclusive for 66 patients (7%, 8%). The experience revealed the needs to complete IOTest™ with a lineage marker (like CD3-APC) and a standardization procedure in detection of fluorescence. Comparative analysis considering 23/44 non-conclusive samples as negative, pointed out a 100% sensitivity and specificity of IOTest™ in determining genetically proved HLA-B27. Specificity of the BD HLA-B27 kit was 94% (two false positives) sticking to the manufacturer instructions but raised to 100% and 94% sensitivity with a cutoff at 8.5 (225 on FACSdiva's scale).

Discussion: The DuraClone B27™ specificity was poor using the manufacturer cutoff but raised to 100% with a 8.0 cutoff instead of 15.

Conclusion: The three flow cytometry kits displayed satisfying performances but need adjustments. The DuraClone B27™ kit seems to be the best while the IOTest™ kit is not conclusive in 8% of cases. In most cases the use of molecular biology can be spared, but genotyping remains essential for patients whose HLA-B27 status cannot be determined with confidence by flow cytometry.

Download full-text PDF

Source
http://dx.doi.org/10.1002/cyto.b.22102DOI Listing

Publication Analysis

Top Keywords

flow cytometry
20
duraclone b27™
12
human leukocyte
8
molecular biology
8
cytometry kits
8
raised 100%
8
hla-b27
6
flow
5
cytometry
5
leukocyte antigen-b27
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!