Background: Previous studies have shown differences in technical image quality between digital breast tomosynthesis (DBT) systems. However, quantitative image quality measurements may not necessarily fully reflect the clinical performance of DBT.

Purpose: To study the subjective image quality of five DBT systems manufactured by Fujifilm, GE, Hologic, Planmed, and Siemens using phantom images.

Material And Methods: A TOR MAM test object with polymethyl methacrylate plates was imaged on five DBT systems from different vendors. Three DBT acquisitions were performed at mean glandular doses of 1.0 mGy, 2.0 mGy, and 3.5 mGy while maintaining a constant phantom set-up. Eight DBT acquisitions with different test plate positions and phantom set-up thicknesses were performed at clinically applied dose levels. Additionally, three conventional two-dimensional mammogram images were acquired with different phantom thicknesses. Six radiologists ranked the systems based on the visibilities of the targets seen in the phantom images.

Results: In the DBT acquisitions performed at comparable dose levels, one system differed significantly from all other systems in microcalcification scores. When using site-specific DBT protocols, significant differences were found between the devices for microcalcification, filament, and low-contrast targets. A strong correlation was observed between the reviewer scores and radiation doses in DBT acquisitions, whereas no such correlation was observed in the 2D acquisitions.

Conclusion: In DBT acquisitions, dose level was found to be a major factor explaining image quality differences between the systems, regardless of other acquisition parameters. Most DBT systems performed equally well at similar dose levels.

Download full-text PDF

Source
http://dx.doi.org/10.1177/02841851221140210DOI Listing

Publication Analysis

Top Keywords

image quality
20
dbt acquisitions
20
dbt systems
16
dose levels
12
dbt
10
breast tomosynthesis
8
systems
8
systems based
8
acquisitions performed
8
mgy mgy
8

Similar Publications

Objective: Epilepsy is a common neurological disease affecting nearly 1% of the global population, and temporal lobe epilepsy (TLE) is the most common type. Patients experience recurrent seizures and chronic cognitive deficits that can impact their quality of life, ability to work, and independence. These cognitive deficits often extend beyond the temporal lobe and are not well understood.

View Article and Find Full Text PDF

Objective: Spinal fusion is a commonly performed surgical procedure used to relieve pain, deformity, and instability of various spinal pathologies. Although there have been attempts to standardize spinal fusion assessment radiologically, there is currently no unified definition that also considers clinical symptomology. This review attempts to create a more holistic and standardized definition of spinal fusion.

View Article and Find Full Text PDF

Objectives: The development of valuable artificial intelligence (AI) tools to assist with ultrasound diagnosis depends on algorithms developed using high-quality data. This study aimed to test the intra- and interobserver agreement of a proposed image-quality scoring system to quantify the quality of gynecological transvaginal ultrasound (TVS) images, which could be used in clinical practice and AI tool development.

Methods: A proposed scoring system to quantify TVS image quality was created following a review of the literature.

View Article and Find Full Text PDF

Purpose: We aimed to determine whether implementation of clinical decision support (CDS) tool integrated into the electronic health record (EHR) of a multi-site academic medical center increased the proportion of patients with American Urological Association (AUA) "high risk" microscopic hematuria (MH) who receive guideline concordant evaluations.

Materials And Methods: We conducted a two-arm cluster randomized quality improvement project in which 202 ambulatory sites from a large health system were randomized to either have their physicians receive at time of test results an automated CDS alert for patients with 'high-risk' MH with associated recommendations for imaging and cystoscopy (intervention) or usual care (control). Primary outcome was met if a patient underwent both imaging and cystoscopy within 180 days from MH result.

View Article and Find Full Text PDF

With the development of the Korean economy, demand for high-quality beef, specifically Hanwoo beef, is escalating, with marbling traits-measured by the widely used marbling score-being a key contributor to meat palatability. The differences between the high-quality and the lower-quality meat, according to the satisfaction of the customers, are not the result from only the degree of marbling but also from the delicacy of the marbling flecks distribution. Using the computer marbling analysis technique, an index for quantifying marbling fineness of 256 sirloin cuts at 12th- 13th thoracic vertebra named F7 index was developed in this study.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!