Background: Remote cardiac monitoring and screening have already become an integral telemedicine component. The wide usage of several different wireless electrocardiography (ECG) devices warrants a validation study on their accuracy and reliability.

Methods: Totally, 300 inpatients with the Nabz Hooshmand-1 handheld ECG device and the GE MAC 1200 ECG system (as the reference) were studied to check the accuracy of the devices in 1 and 6-limb lead performance. Simultaneous 10-second resting ECGs were assessed for the most common ECG parameters in lead I. Afterward, 6-lead ECGs (limb leads), were performed immediately and studied for their morphologies.

Results: Of the 300 patients, 297 had acceptable ECG quality in both devices for simultaneous lead I ECGs. The ECGs were inspected on-screen by a cardiologist for their rhythms, rates, axes, numbers, morphologies of premature atrial and ventricular beats, morphologies and amplitudes of PQRST waves, P-wave durations, QRS-wave durations, P-R intervals, and QT intervals. No significant differences were detected between the devices, and no major abnormalities were missed. Six-limb lead ECGs were obtained in 284 patients, of whom 281 had acceptable quality in ECGs by both devices. The morphology matching evaluation of the ECGs demonstrated an overall 98% compatibility rate, with the highest compatibility in lead I and the lowest in lead augmented vector foot.

Conclusions: The diagnosis of critical pathological rhythms, including atrial fibrillation and high-grade atrioventricular node block, was not missed by the Nabz Hooshmand-1 and GE MAC 1200 ECG devices. Accordingly, rhythm detection as the primary purpose of handheld ECG devices was highly accurate. Both devices had acceptable sensitivity to diagnose long P-R and long and short QT intervals. Although the modern technology of smartphones and the physical inability for the 6-limb mode might cause old patients difficulty in utilizing such devices, their use for screening and follow-up is safe.

Download full-text PDF

Source
http://dx.doi.org/10.1097/HPC.0000000000000303DOI Listing

Publication Analysis

Top Keywords

handheld ecg
12
ecg devices
12
devices
9
ecg
8
ecg device
8
validation study
8
nabz hooshmand-1
8
mac 1200
8
1200 ecg
8
lead ecgs
8

Similar Publications

Introduction: Traditional monitoring of athletes with cardiac symptoms is limited due to sport-specific considerations and the intermittent nature of symptoms. Some portable electrocardiogram (ECG) devices may have more diagnostic utility than traditional monitoring. Their accuracy, advantages, and limitations should be considered when a clinician is considering the most appropriate device for investigation of an athlete's symptoms.

View Article and Find Full Text PDF

Self-recorded, single-lead electrocardiograms (ECGs) are increasingly used to diagnose arrhythmias. However, they can be of variable quality, affecting the reliability of interpretation. In this analysis of ECGs collected in atrial fibrillation screening studies, our aims were to: (i) determine the quality of ECGs when recorded unsupervised; and (ii) investigate whether telephone training improved ECG quality.

View Article and Find Full Text PDF

On the detection of acute coronary occlusion with the miniECG.

Eur Heart J Digit Health

November 2024

Department of Cardiology, University Medical Center Utrecht, Utrecht University, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.

View Article and Find Full Text PDF

Objective: To examine the feasibility of implementing remote atrial fibrillation (AF) self-screening among older people supported by a remote central monitoring system.

Design: Process evaluation of the Mass AF randomised clinical trial (ACTRN12621000184875) with one-to-one semistructured interviews using interview guides underpinned by the Critical Realism approach and coded using the UK Medical Research Council Guidance of Process Evaluation Framework.

Setting And Participants: Community-dwelling people aged ≥75 years from both genders (ratio 1:1) and urban/rural (ratio 2:1) in Australia.

View Article and Find Full Text PDF
Article Synopsis
  • Hyperkalemia, a condition of elevated potassium levels, is linked to higher mortality rates in patients in cardiac intensive care units (CICU), and an AI-powered electrocardiogram (ECG) can predict this condition effectively.
  • The study involved over 11,000 CICU patients and found that AI-ECG could identify hyperkalemia in patients even when laboratory tests showed normal potassium levels, with a notable percentage predicted to have the condition.
  • Results indicated that patients identified by AI-ECG as hyperkalemic faced increased in-hospital mortality and reduced 1-year survival, suggesting AI-ECG offers valuable risk assessment beyond conventional lab measurements.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!