Background And Purpose: Metallic implants cause artefacts in computed tomography (CT) images and can introduce significant errors to structure visualisation and dosimetric calculation within the radiotherapy planning process. This study evaluated an orthopaedic metal artefact reduction algorithm and its effect on the CT number, image noise, structure delineation, and treatment dose.

Methods: Raw CT data were reconstructed using standard filtered back projection and an artefact reduction algorithm to create 'standard' and 'corrected' images. A phantom containing tissue-mimicking inserts and two titanium plugs was imaged. The average CT number was compared to baseline data acquired without metal inserts. Data from 11 pelvic external beam radiotherapy (EBRT) patients with bi- or uni-lateral hip implants were retrospectively analysed. The clinically used treatment plans were re-computed on the corrected images. A prostate-mimicking phantom containing metal 'implants' was imaged, and 11 observers contoured both reconstructions.

Results: The artefact reduction algorithm improved the CT number in those areas most affected by metal artefacts and decreased noise by 19 % (P =.04) Changes in dose distributions on corrected images compared to those calculated using the current clinical protocol were clinically insignificant. Volumes contoured on the corrected phantom images had larger Dice coefficients than those contoured on the standard images (P =.001), as well as a 36 % lower standard deviation in volumes.

Conclusion: This study demonstrates that the metal artefact reduction software reduces the error in CT numbers, can improve delineation accuracy, and can reduce inter-observer variability. It has the potential to streamline the planning pathway and improve treatment planning accuracy.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9674537PMC
http://dx.doi.org/10.1016/j.phro.2022.11.007DOI Listing

Publication Analysis

Top Keywords

artefact reduction
20
reduction algorithm
16
metal artefact
12
treatment planning
8
corrected images
8
images
6
artefact
5
reduction
5
metal
5
impact metal
4

Similar Publications

Homonuclear decoupled INADEQUATE NMR methods with improved sensitivity and resolution in solid-state NMR.

J Magn Reson

December 2024

Univ. Lille, CNRS, Centrale Lille, Univ. Artois, UMR 8181, UCCS - Unité de Catalyse et Chimie du Solide, F-59000 Lille, France. Electronic address:

The two-dimensional (2D) refocused INADEQUATE NMR experiment, which correlates double-quantum (DQ) and single-quantum (SQ) coherences, is widely used to probe the chemical connectivities in solids. Nevertheless, the multiplets along the F dimension reduce the resolution and sensitivity of this experiment. The Composite-Refocusing (CR) technique with two excitation pulses has been proposed to suppress these multiplets in 2D INADEQUATE spectra of liquids.

View Article and Find Full Text PDF

LC-ESI-MS/MS is a preferred method for detecting and identifying metabolites, including those that are unpredictable from the genome, especially in basal metazoans like Cnidaria, which diverged earlier than bilaterians and whose metabolism is poorly understood. However, the unexpected appearance of a "ghost peak" for dopamine, which exhibited the same m/z value and MS/MS product ion spectrum during an analysis of Nematostella vectensis, a model cnidarian, complicated its accurate identification. Understanding the mechanism by which "ghost peaks" appear is crucial to accurately identify the monoamine repertoire in early animals so as to avoid misassignments.

View Article and Find Full Text PDF

DDoCT: Morphology preserved dual-domain joint optimization for fast sparse-view low-dose CT imaging.

Med Image Anal

December 2024

School of Physics, Beihang University, Beijing, China; Hangzhou International Innovation Institute, Beihang University, Hangzhou, China; Tianmushan Laboratory, Hangzhou, China. Electronic address:

Computed tomography (CT) is continuously becoming a valuable diagnostic technique in clinical practice. However, the radiation dose exposure in the CT scanning process is a public health concern. Within medical diagnoses, mitigating the radiation risk to patients can be achieved by reducing the radiation dose through adjustments in tube current and/or the number of projections.

View Article and Find Full Text PDF

Surface electromyography () is useful for studying muscle function and controlling prosthetics, but crosstalk from nearby muscles often limits its effectiveness. High-density surface EMG () improves spatial resolution, allowing for the isolation of in the densely packed forearm muscles. This study assessed for localizing and evaluated the impact of spatial filters on crosstalk reduction.

View Article and Find Full Text PDF

Reduction of artifacts associated with missing data in coherent diffractive imaging.

J Synchrotron Radiat

January 2025

ESRF - The European Synchrotron, 71 Avenue des Martyrs, 38000 Grenoble, France.

Coherent diffractive imaging experiments often collect incomplete datasets containing regions that lack any measurements. These regions can arise because of beamstops, gaps between detectors, or, in tomography experiments, a missing wedge of data due to a limited sample rotation range. We describe practical and effective approaches to mitigate reconstruction artifacts by bringing uniqueness back to the phase retrieval problem.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!